2  OPERATING INSTRUCTIONS 

 

 

 

2.1  WITH GENTEC-EO MONITORS 

 

To  operate  the  PH  and  PE  Series  photo  detector,  connect  the  detector  head to the input socket on 
Gentec-eo  monitors  with  the  cable  supplied  by  Gentec-EO  before  turning  the  monitor  on.    (See 
monitor

’s instruction manual.) 

 
Adjust  the  reading  to  zero  before  measurements  for  best  accuracy.    See  procedure  in  section  2.2 
below.   
 
If  you  are  using  the  fiber  optic  adapter,  put  the  black  protective  cover  on  the  detector  head  before 
making the offset nulling. 
 
If you want to subtract the room light do the offset nulling with the cover off. 
   

2.2  QUICK POWER MEASUREMENT PROCEDURE 

This section will show you the fastest way of making a laser power measurement with Gentec-eo monitors 
and power detector.   

The monitor automatically recognizes all Gentec-EO photo detectors heads of version 5 or higher.  Then 
the monitor automatically downloades all custom technical data for the detector from the EEPROM in the 
DB-15  connector.    This  data  includes  sensitivity,  model,  serial  number,  version,  wavelength  correction 
factors, and time response. 

Quick power measurement procedure: 

 

 

 

 

 

 

 

 

 

 

Adjusting the zero (steps 8 to 10)

 

 

Содержание PE10B-Si

Страница 1: ...to obtain a return authorization number Return the material to the appropriate address below Contacting Gentec Electro Optics Inc To help us answer your calls more efficiently please have the model n...

Страница 2: ...idity or surface contaminants exposure to flame solvents or water and connection to improper electrical voltage TABLE OF CONTENTS WARRANTY I First Year Warranty i Lifetime Warranty i TABLE OF CONTENTS...

Страница 3: ...TO DETECTOR DIMENSIONS 3 FIG 1 2 TYPICAL GE TEMPERATURE DEPENDENCE VS THE WAVELENGTH 10 FIG 1 3 TYPICAL SI AND SIUV TEMPERATURE DEPENDENCE VS THE WAVELENGTH 10 FIG 1 4 PH100 SI TYPICAL SPECTRAL RESPON...

Страница 4: ...uoted maximum ranges are average power at the 1064 nm wavelength for PH100 Si and PH20 Ge 532 nm for PH100 SiUV and 850 nm for PH100 SiUV with OD 2 quoted minimum ranges are average power at the 980nm...

Страница 5: ...ific PH and PE Series head Wavelengths are not available where the sensor physical properties don t allow it The Gentec eo monitors use the data in this connector to adjust their characteristics autom...

Страница 6: ...Version 2 0 PH Series Instruction Manual Gentec Electro Optics Inc All rights reserved 3 1 2 2 Dimensions FIG 1 1 PHOTO DETECTOR DIMENSIONS...

Страница 7: ...30 mW 320 mW cm2 1064 nm Minimum Powera 0 6 nW 980 nm 0 6 nW 850 nm 2 nW 1550 nm Damage threshold Maximum Average Power Density 100 W cm 2 100 W cm 2 100 W cm 2 Typical Detector saturation currentb 6...

Страница 8: ...W 980 nm 0 45 A W 850 nm 0 98 A W 1550 nm With attenuator PH100 Si OD1 OD2 PH100 SiUV OD1 OD2 PH20 Ge OD1 OD2 Maximum Power typ w OD 1 300 mW 1064 nm 25 mW 532 nm 300 mW 1064 nm Minimum Power a w OD 1...

Страница 9: ...measurable power M Link 200 W S Link 175 W Maestro 150 W 633nm M Link 32 W S Link 26 W Maestro 22 W 1047 nm Maximum Average Power Density 65 mW cm2 532 nm 320 mW cm2 1064 nm Noise equivalent power 10...

Страница 10: ...Repetition Rate 1000 Hz Max Pulse Width 10 s Maximum Measurable Energy M Link 0 16 J S Link 0 13 J Maestro 0 11 J 633nm M Link 3 2 nJ S Link 2 6 nJ Maestro 2 4 nJ 1047 nm Maximum Energy Density 5 J c...

Страница 11: ...ack protective cover on the detector head before making the offset nulling If you want to subtract the room light do the offset nulling with the cover off 2 2 QUICK POWER MEASUREMENT PROCEDURE This se...

Страница 12: ...component not related to the laser power This will add a systematic error to absolute power measurements This error may disappear from relative power measurements When you subtract two measurements m...

Страница 13: ...0 PH Series Instruction Manual Gentec Electro Optics Inc All rights reserved 10 FIG 1 2 TYPICAL GE TEMPERATURE DEPENDENCE VS THE WAVELENGTH FIG 1 3 TYPICAL SI AND SIUV TEMPERATURE DEPENDENCE VS THE W...

Страница 14: ...saturate the detector To know if the detector is saturated use procedure 4 3 1 Be careful when making the attenuator calibration in pulse mode The peak power must be in the linear region of the photo...

Страница 15: ...esired Wavelength The difference between the desired wavelength and the inferior wavelength 4 5 1 Example You have a PH100 Si and your laser is at 632 8 nm See your CERTIFICATE OF CALIBRATION for the...

Страница 16: ...Version 2 0 PH Series Instruction Manual Gentec Electro Optics Inc All rights reserved 13 FIG 1 4 PH100 SI TYPICAL SPECTRAL RESPONSE FIG 1 5 PH100 SIUV PH10B SI PE10B SI TYPICAL SPECTRAL RESPONSE...

Страница 17: ...Version 2 0 PH Series Instruction Manual Gentec Electro Optics Inc All rights reserved 14 FIG 1 6 PH20 GE PH5B GE PE5B GE TYPICAL SPECTRAL RESPONSE...

Страница 18: ...center when the detector reaches its end of life Breaking the calibration seal or opening the monitor will void the detector warranty The complete Detector contains 1 Detector with wires or DB 15 1 i...

Страница 19: ...Version 2 0 PH Series Instruction Manual Gentec Electro Optics Inc All rights reserved 16...

Отзывы: