20
4. Troubleshooting and FAQs
Error messages
Error Message
Possible causes
Solutions
Buffer Error
Can’t read cassette chip/no
cassette
Replace cassette/ insert
cassette
No Tank
Cassette not secured in
position
Check Pins and Reinsert
cassette securely
Pump Error
Pump malfunction
Contact GenScript
System Error
System malfunction
Contact GenScript
FAQ
Description
Possible causes
Solutions
Strong background
Insufficient wash
1. Increase number
of wash cycles (set
to 5 cycles for 3
minutes each)
2. Increase pump
flow per wash
cycles (set to
normal)
Weak signal
1. Antibody is too
diluted
2. Instrument needs
cleaning
3. Insufficient
antibody
incubation time
1. Reduce antibody
dilution factor
2. Clean eZwest Lite
instrument
3. Increase antibody
incubation time or
reduce incubation
pump flow
Fluorescent detects
only strong signal
around spacer
Weak target signal
contrasted out by
interference
1. Adjust detection
area to omit
spacer position
2. Cover spacer
signal region with
paper or tape