8.   Place the gel on the membrane. Gently remove air bubbles between the 

      gel and the membrane.

9.   Select appropriate Gel Window according to actual size of the gel (see 

      the table below). Place Gel Window on the gel. Ensure that the Gel 

      Window fully covers the margin of the membrane.

10.   Tear the sealing of the 1×

e

Blot

TM

 Cathode Pad package. Remove the 

        

e

Blot

TM

 Cathode Pad from the package and place it on the gel.

Note:  Gel Window is used as the spacer between anode pad and cathode pad to 

            prevent short circuit.

Note:  During assembling of the transfer stack, make sure to remove all the air bubbles

           trapped between the transfer pads, pre-run gel and the membrane, which may

           prevent the transfer of proteins and cause empty spots on the transferred membrane.

6.   Place the equilibrated membrane on the 

e

Blot

TM

 Anode Pad. Gently remove 

      air bubbles between the membrane and the anode pad using the small

     shovel supplied with the device.

7.   Carefully remove the pre-run gel containing your protein samples from the 

      gel cassette and briefly rinse the gel with distilled water in another shallow 

      tray to remove any small gel pieces attached to the gel and facilitate easy 

      positioning of the gel on the membrane. 

Pre-run gel

Gel with size > 80 × 60 mm

Gel with size 66 × 56 mm to 80 × 60 mm

Gel with size < 66 × 56 mm

Gel Window

Gel Window 88 × 78 mm

Gel Window 76 × 64 mm

Gel Window 66 × 56 mm

Содержание eBlot

Страница 1: ...iscataway NJ 08854 Tel 732 885 9188 732 885 9688 Fax 732 210 0262 732 885 5878 Email product genscript com Version No 08152014 For fast semi dry electroblotting of proteins from mini polyacrylamide ge...

Страница 2: ...ions eBlotTM Protein Transfer Device Accessory Products Introduction Quick Reference Guide Protocols Troubleshooting Examples of Results Technical Support Warranty Appendix 1 3 5 7 8 11 12 21 22 23 24...

Страница 3: ...mponent eBlotTM Protein Transfer Pad Nitrocellulose membranes 20 sheets PVDF membranes 20 sheets eBlotTM Equilibration Buffer 125 ml Gel Window 66 mm 56 mm Gel Window 76 mm 64 mm Gel Window 88 mm 78 m...

Страница 4: ...ely from GenScript The specifications for eBlotTM Graphite Electrode are listed below eBlotTM Protein Transfer Pads Specifications eBlotTM Anode Pad eBlotTM Cathode Pad Materials 90 mm l 80 mm w 2 5 m...

Страница 5: ...both left and right status lights are on the device is working at numbering mode The Reset button is used to clear parameters The Min button is used to shift between transferring and numbering mode a...

Страница 6: ...s Equilibration Buffer and Gel Window Each pack of eBlotTM Protein Transfer Pad contains an 1 eBlotTM Cathode Pad and an 1 eBlotTM Anode Pad presoaked with proprietary cathode buffer and anode buffer...

Страница 7: ...etary fast semi dry blotting technique developed by GenScript To use the eBlotTM Protein Transfer System for rapid protein blotting assemble the eBlotTM Protein Transfer Pad with your pre run gel and...

Страница 8: ...ng the number of uses of the graphite anode Replacement of worn graphite anode Action Connect eBlotTM Device to an electrical outlet and power switch is on Assemble transfer stack on the device and cl...

Страница 9: ...e Protocols continued Protocols continued 8 Attach the power cord to the AC inlet and then to the electrical outlet Use only properly grounded AC outlets and power cords 9 When the electrophoresis of...

Страница 10: ...nd cathode pad to prevent short circuit Note During assembling of the transfer stack make sure to remove all the air bubbles trapped between the transfer pads pre run gel and the membrane which may pr...

Страница 11: ...o perform the replacement Note After about 40 runs take out the graphite electrode soak in distilled water for 30 minutes dry the surface with a clean paper towel then leave air dry overnight with gra...

Страница 12: ...phite Electrode out of the device 3 Tear the sealing of a new Graphite Electrode package and take the new Graphite Electrode out of the package Place the new Graphite Electrode into the anode tank and...

Страница 13: ...ansfer stack is assembled correctly use the eBlotTM Anode Pad first followed by the membrane the pre run gel Gel Window and eBlotTM Cathode Pad Check the transfer stack and ensure Gel Window covered c...

Страница 14: ...vent or sample For any inquiry or request for repair service contact GenScript after confirming the model and serial number of your instrument For your protection items being returned must be insured...

Страница 15: ...26 25 Appendix Appendix continued eBlotTM Protein Transfer System www genscript com eBlotTM Protein Transfer System www genscript com...

Страница 16: ...27 Appendix continued eBlotTM Protein Transfer System www genscript com...

Отзывы: