6
2.2 Connection of Power Supply
1.AC power must have the grounding protection.
2.Do not connect the ground wire of indicator directly to
the GND of other equipments.
2.3 Connection of Load Cell
Please refer to the picture below to connect load cells to Tuxon-A. When you use 4-
wired load cells, you must bridge the SN+ with EX+ and bridge the SN- with EX-.
The signal definition of each port of the load cell connector is as follows:
Port
EX+
SN+
EX-
SN-
SIG+
SIG-
SHLD
Definition
Exc
Sense+
Excitation-
Sense-
Signal- Shield
2.3.1 6-wired Connection