GM8802F-2 Field Mounting Transmitter
GM8802-F2110101
24
0404
0403
Reserve
Write operation
:
Ineffective, and
return error message.
Read operation
:
read fixed
0
0405
0404
Channel no.
1
zeroing
Write
:
ON
for zero calibration
OFF
for no operation.
Read
:
1
for zero status
0
for nonezero status
0406
0405
Channel no.
2
zeroing
Write
:
ON
for zero calibration
OFF
for no operation.
Read
:
1
for zero status
0
for nonezero status
0407
0406
Reserve
Write operation
:
Ineffective, and
return error message.
Read operation
:
read fixed
0
0408
0407
Reserve
Write operation
:
Ineffective, and
return error message.
Read operation
:
read fixed
0
0409
0408
Channel no.
1
start / shut
Write
:
ON
to start
AD
OFF
to shut
AD
Read
:
0
means
AD
shut
1
means
AD
start
0410
0409
Channel no.
2
start / shut
Write
:
ON
to start
AD
OFF
to shut
AD
Read
:
0
means
AD
shut
1
means
AD
start
0411
0410
Reserve
Write operation
:
Ineffective, and
return error message.
Read operation
:
read fixed
0
0412
0411
Reserve
Write operation
:
Ineffective, and
return error message.
Read operation
:
read fixed
0
0413
0412
System restart
Write
:
ON
restart system
OFF
for no operation
Read
:
0
0414
0413
Restore default setting
Write
:
ON
return to default setting
OFF
for no operation
Read
:
0
0415
0414
Modbus
high / low byte
mode
Write
:
ON
for low byte mode
OFF
for high byte mode
Read
:
0
for high byte mode
1
for low byte mode
Note: Fix 4 decimal digits when read millivolt.