GM8802C-D Weighing Transmitter
GM8802C-D110101 49
Received data sequence is incorrect:
STX
Scale ID
C
M
N
O
CRC
CR
LF
Here:
M
——
1bit, 4DH
Division value
——
2bits,
writing Min. division value
(
1
、
2
、
5
、
10
、
20
、
50
)
Max. Capacity
——
6bits,
writing Max. Capacity
For example:
Command from upper computer
:
02 30 31 43 4D 30 35 30 31 34 30 30 30 33 37 0D 0A
(
Writing data
:
division value is
5
,
Max. capacity is
14000
)
Correct response format
:
02 30 31 43 4D 4F 4B 39 37 0D 0A
Incorrect response format
:
02 30 31 43 4D 4E 4F 30 30 0D 0A
3 . Calibration the Sensitivity of Sensor
Sending Command format:
STX
Scale ID
C
E
Sensitivity of Sensor
CRC
CR
LF
Received data sequence is correct:
STX
Scale ID
C
E
O
K
CRC
CR
LF
Received data sequence is incorrect: