GM8802C-DWeighing Transmitter
44 GM8802C-D110101
5.3.2.3 Reading the Sensitivity of Sensor
Sending Command format:
STX
Scale ID
R
E
CRC
CR
LF
Received data sequence is correct:
STX
Scale ID
R
E
DDDDDD
CRC
CR
LF
Received data sequence is incorrect:
STX
Scale ID
R
E
N
O
CRC
CR
LF
Here:
E
——
1bit
,
45H
DDDDDD
——
6bits, 000000~000001
corresponding
2mV/V
、
3mV/V
.
For example:
Command from upper computer
:
02 30 31 52 45 35 30 0D 0A
Correct response format
:
02 30 31 52 45 30 30 30 30 30 30 33 38 0D 0A
(the sensitivity of
sensor is
2mV/V )
Incorrect response format
:
02 30 31 52 45 4E 4F 30 37 0D 0A