SW2 - Setting for X-Ray Output
032-0207-EN Rev 0
4-2
Switch Settings
SW2 - Setting for X-Ray Output
Normally the dip switch setting for x-ray output, SW2 position 4, should not
be changed. If for some reason the setting has been changed, the x-ray
output will be disabled. All other functions will appear normal.
Note!
It is recommended that this procedure be performed only by a qualified
Gendex service agent.
WARNING
Ensure that you follow the procedure referenced in the following step.
This equipment must always be electrically disconnected from the
mains electrical supply (external power source) before beginning any
procedure.
1. Follow the steps outlined in
Opening the Master Control
, beginning on
page 1-2, to open the Master Control.
2. Locate the series of dip switches labeled SW2 located at the bottom left
of the Logic Board.
3. Set the position 4 switch to the OFF position. All other switches should
remain in the OFF position.
Figure 4-1
Setting the SW2 Dip Switch
4. Follow the steps outlined in
Closing the Master Control
on page 1-8 to
close the Master Control.