System Functions
WARNING
Before starting any procedure, review the safety information provided
in the Gendex expert
®
DC User Manual.
If problems are found, refer to the appropriate section in this Service Manual
for:
•
Adjustment procedures for the Horizontal Arm, Articulated Arm, and
Tubehead (page 3-1)
•
Maintenance procedure for the Articulated Arm (page 3-11)
•
Troubleshooting information (page 5-1)
•
Electrical block diagrams (page 5-20).
IMPORTANT!
Contact GENDEX Technical Support if additional information is required.
____ 1.
Tubehead
- Check for oil leaks or other evidence that could
indicate internal damage. Replace the Tubehead if necessary.
____ 2.
Tubehead Rotation
- Ensure that the Tubehead maintains its
position around the horizontal axis while remaining easy to rotate
and position. For horizontal dropping, refer to page 3-2 for the
procedure.
Also check for unwanted pivoting of the Tubehead around the
vertical axis. Problems with vertical pivot are typically caused
when the Horizontal Arm is not properly leveled although the
problem can simply be caused when the cabling inside the
Tubehead Yoke is too tight.
____ 3.
Mounting
- Be sure that the wall support is adequate and that the
system is properly mounted to the wall.
____ 4.
Power Switch
- Verify that the switch is working properly and
that the Ready Indicator Lamp is illuminated when the power
switch is in the ON position.