Chapter 10 - Care and Maintenance
P R E L I M I N A R Y
These tests may refer to specific safety analyzer equipment as an example. Always refer to the safety
analyzer’s user manual that will be used to perform the tests.
•
Check for missing or loose enclosure covers that could allow access to internal live parts.
•
Examine the mains cord, mains plug and appliance inlet for damaged insulation and adequacy of
strain relief and cable clamps.
•
Locate and examine all associated transducers. Inspect the cables and strain relief at each end.
Inspect the transducer enclosure and lens for cracks, holes and similar defects.
Prior to initiating any electrical test, the Ultrasound system must be visually inspected. Perform the
following visual checks:
Equipment users must ensure that safety inspections are performed whenever damage is
suspected and on a regular basis in accordance with local authorities and facility procedures.
Do not use the Ultrasound system or individual probes which fail any portion of the safety test.