Revision A
GE T2100 Treadmill
6-19
2021403-031
Troubleshooting: TDU Troubleshooting Tables and Status LEDs
0x05
Fatal
(Halts
Current
Operation)
Communication Error:
!
<01> Could not check for drive
errors.
!
<02> Could not check for drive
axis errors.
!
<03> Could not read the belt
speed from the drive.
!
<11> Invalid error bit pattern
while checking for drive errors.
!
<12> Invalid error bit pattern
while checking for drive axis
errors.
!
<13> Invalid speed detected
while reading belt speed from
drive.
!
<255> Command not received
from host within 4 seconds
while the belt was enabled.
Communication error between the processor board and the drive
controller, or the processor board and the host.
1. Powercycle the GE T2100 Treadmill and monitor the GE T2100
Treadmill for reoccurrence of this error. (This may be an isolated
incident.)
2. Check for loose wiring and connections between the drive
controller and the e-box.
3. Check for loose wiring and connections between the treadmill
and the host.
4. Check the terminal connections on the X1 connector on the
“Drive Controller X1 PIN descriptions”
5. Check the power board and processor board test points and
verify they are operating within the test range. See
Supply Voltage Checks and Fuses”
6. Replace the drive controller with a known working drive
controller. (If the problem persists...)
7. Replace the processor board. See
0x06
Fatal
(Halts
Current
Operation)
Over-voltage
Measures power into the GE T2100 Treadmill.
1. Powercycle the GE T2100 Treadmill and monitor the GE T2100
Treadmill for reoccurrence of this error. (This may be an isolated
incident.)
2. Check the fuses. See
29.
3. Check the power board and processor board test points and
verify they are operating within the test range. See
Supply Voltage Checks and Fuses”
4. Replace the power board. See
0x07
Fatal
(Halts
Current
Operation)
Under-voltage:
!
<00> Under-voltage detected
while system still operating.
!
<255> Under-voltage caused
system to shut down while the
belt was enabled, or while the
elevation motor was active.
Measures power into the GE T2100 Treadmill.
1. Powercycle the GE T2100 Treadmill and monitor the GE T2100
Treadmill for reoccurrence of this error. (This error logs each
time the treadmill is powered off. This may be an isolated
incident.)
2. Check the fuses. See
29.
3. Check the power board and processor board test points and
verify they are operating within the test range. See
Supply Voltage Checks and Fuses”
4. Monitor AC power into the GE T2100 Treadmill. Watch for
inadequate power supply.
5. Replace the power board. See
on page 7-28. (If the problem persists...)
6. Replace the processor board. See
0x08
Warning
Belt Start
Not an error. Indicates the belt was started at this time.
TDU Event /
Error Code
Status
Definition and Detail
Description
Check These Items
Содержание T2100
Страница 1: ...GE T2100 Treadmill Service Manual 2021403 031 Revision A ...
Страница 8: ...vi GE T2100 Treadmill Revision A 2021403 031 ...
Страница 9: ...Revision A GE T2100 Treadmill 1 1 2021403 031 1 Introduction ...
Страница 10: ...1 2 GE T2100 Treadmill Revision A 2021403 031 For your notes ...
Страница 19: ...Revision A GE T2100 Treadmill 2 1 2021403 031 2 Equipment Overview ...
Страница 20: ...2 2 GE T2100 Treadmill Revision A 2021403 031 For your notes ...
Страница 26: ...2 8 GE T2100 Treadmill Revision A 2021403 031 Equipment Overview General Description For your notes ...
Страница 27: ...Revision A GE T2100 Treadmill 3 1 2021403 031 3 Installation ...
Страница 28: ...3 2 GE T2100 Treadmill Revision A 2021403 031 For your notes ...
Страница 44: ...3 18 GE T2100 Treadmill Revision A 2021403 031 Installation Functional Checkout For your notes ...
Страница 45: ...Revision A GE T2100 Treadmill 4 1 2021403 031 4 Maintenance ...
Страница 46: ...4 2 GE T2100 Treadmill Revision A 2021403 031 For your notes ...
Страница 51: ...Revision A GE T2100 Treadmill 5 1 2021403 031 5 Theory of Operation ...
Страница 52: ...5 2 GE T2100 Treadmill Revision A 2021403 031 For your notes ...
Страница 72: ...5 22 GE T2100 Treadmill Revision A 2021403 031 Theory of Operation Electrical Requirements For your notes ...
Страница 73: ...Revision A GE T2100 Treadmill 6 1 2021403 031 6 Troubleshooting ...
Страница 74: ...6 2 GE T2100 Treadmill Revision A 2021403 031 For your notes ...
Страница 102: ...6 30 GE T2100 Treadmill Revision A 2021403 031 Troubleshooting Power Supply Voltage Checks and Fuses For your notes ...
Страница 103: ...Revision A GE T2100 Treadmill 7 1 2021403 031 7 Assembly Disassembly ...
Страница 104: ...7 2 GE T2100 Treadmill Revision A 2021403 031 For your notes ...
Страница 132: ...7 30 GE T2100 Treadmill Revision A 2021403 031 Assembly Disassembly PCBs and Power Supply Replacement For your notes ...
Страница 133: ...Revision A GE T2100 Treadmill 8 1 2021403 031 8 Ordering Parts ...
Страница 134: ...8 2 GE T2100 Treadmill Revision A 2021403 031 For your notes ...
Страница 137: ...Revision A GE T2100 Treadmill A 1 2021403 031 A Appendix A Technical Specifications ...
Страница 138: ...A 2 GE T2100 Treadmill Revision A 2021403 031 For your notes ...
Страница 142: ...A 6 GE T2100 Treadmill Revision A 2021403 031 Appendix A Technical Specifications Technical Specifications ...
Страница 143: ...Revision A GE T2100 Treadmill B 1 2021403 031 B Appendix B Electromagnetic Compatibility EMC ...
Страница 144: ...B 2 GE T2100 Treadmill Revision A 2021403 031 For your notes ...