B-6
GE T2100 Treadmill
Revision A
2021403-031
Appendix B – Electromagnetic Compatibility (EMC): Electromagnetic Immunity
Electromagnetic Immunity
Guidance and manufacturer’s declaration – electromagnetic immunity
The
GE
T2100 Treadmill
is intended for use in the electromagnetic environment specified below. The customer or user of the
GE
T2100 Treadmill
should assure that it is used in such an environment.
Immunity Test
Compliance
Test level
Compliance
Level
Electromagnetic Environment – Guidance
Portable and mobile RF communications equipment should
be used on closer to any part of the [equipment or system]
,
including cables, than the recommended separation distance
calculated fro the equation applicable to the frequency of the
transmitter.
Recommended separation distance
Conducted RF
IEC 60601-1-2:2004
o
EN 61000-4-
6:1996 / A1:2001
3 Vrms
150 KHz to 80 MHz
@ 2 Hz mod.
3 V rms
d
= [3.5/V1]
Radiated RF
IEC 60601-1-2:2004
o
EN 61000-4-
3:2002 / A1:2002
3 V/m
80 MHz to 2.5 GHz
@ 1 Khz mod.
3 V/m
d
= [3.5/E1]
80 MHz to 800 MHz
d
= [7/E1]
800 MHz to 2.5 GHz
where
P
is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and
d
is the recommended separation distance
in meters (m).
Field strengths from fixed RF transmitters, as determined by
an electromagnetic site survey
a
, should be less than the
compliance level in each frequency range.
Interference may occur in the vicinity of equipment marked
with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by reflection from structures, objects, and
people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radio, AM and
FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due
to fixed RF transmitters, and electromagnetic site survey should be considered. If the measured field strength in the location in which
the
GE T2100 Treadmill
is used exceeds the applicable RF compliance level above, the
GE T2100 Treadmill
should be observed
to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or
relocating the
GE T2100 Treadmill
.
b
Over the frequency range 150 KHz to 80 MHz, field strengths should be less than
3
V/m.
P
P
P
Содержание T2100
Страница 1: ...GE T2100 Treadmill Service Manual 2021403 031 Revision A ...
Страница 8: ...vi GE T2100 Treadmill Revision A 2021403 031 ...
Страница 9: ...Revision A GE T2100 Treadmill 1 1 2021403 031 1 Introduction ...
Страница 10: ...1 2 GE T2100 Treadmill Revision A 2021403 031 For your notes ...
Страница 19: ...Revision A GE T2100 Treadmill 2 1 2021403 031 2 Equipment Overview ...
Страница 20: ...2 2 GE T2100 Treadmill Revision A 2021403 031 For your notes ...
Страница 26: ...2 8 GE T2100 Treadmill Revision A 2021403 031 Equipment Overview General Description For your notes ...
Страница 27: ...Revision A GE T2100 Treadmill 3 1 2021403 031 3 Installation ...
Страница 28: ...3 2 GE T2100 Treadmill Revision A 2021403 031 For your notes ...
Страница 44: ...3 18 GE T2100 Treadmill Revision A 2021403 031 Installation Functional Checkout For your notes ...
Страница 45: ...Revision A GE T2100 Treadmill 4 1 2021403 031 4 Maintenance ...
Страница 46: ...4 2 GE T2100 Treadmill Revision A 2021403 031 For your notes ...
Страница 51: ...Revision A GE T2100 Treadmill 5 1 2021403 031 5 Theory of Operation ...
Страница 52: ...5 2 GE T2100 Treadmill Revision A 2021403 031 For your notes ...
Страница 72: ...5 22 GE T2100 Treadmill Revision A 2021403 031 Theory of Operation Electrical Requirements For your notes ...
Страница 73: ...Revision A GE T2100 Treadmill 6 1 2021403 031 6 Troubleshooting ...
Страница 74: ...6 2 GE T2100 Treadmill Revision A 2021403 031 For your notes ...
Страница 102: ...6 30 GE T2100 Treadmill Revision A 2021403 031 Troubleshooting Power Supply Voltage Checks and Fuses For your notes ...
Страница 103: ...Revision A GE T2100 Treadmill 7 1 2021403 031 7 Assembly Disassembly ...
Страница 104: ...7 2 GE T2100 Treadmill Revision A 2021403 031 For your notes ...
Страница 132: ...7 30 GE T2100 Treadmill Revision A 2021403 031 Assembly Disassembly PCBs and Power Supply Replacement For your notes ...
Страница 133: ...Revision A GE T2100 Treadmill 8 1 2021403 031 8 Ordering Parts ...
Страница 134: ...8 2 GE T2100 Treadmill Revision A 2021403 031 For your notes ...
Страница 137: ...Revision A GE T2100 Treadmill A 1 2021403 031 A Appendix A Technical Specifications ...
Страница 138: ...A 2 GE T2100 Treadmill Revision A 2021403 031 For your notes ...
Страница 142: ...A 6 GE T2100 Treadmill Revision A 2021403 031 Appendix A Technical Specifications Technical Specifications ...
Страница 143: ...Revision A GE T2100 Treadmill B 1 2021403 031 B Appendix B Electromagnetic Compatibility EMC ...
Страница 144: ...B 2 GE T2100 Treadmill Revision A 2021403 031 For your notes ...