Guidance and manufacturer’s declaration – electromagnetic immunity
The
GE
T2100 Treadmill
is intended for use in the electromagnetic environment specified below. The customer or user of the
GE
T2100 Treadmill
should assure that it is used in such an environment.
Immunity Test
Compliance
Test level
Compliance level
Electromagnetic Environment – Guidance
Electrostatic discharge
(ESD)
IEC 60601-1-2:2004
o
EN 61000-4-
2:1995 / A1:1998 /
A2:2001
± 2/4/6 kV indirect
± 2/4/6 kV direct
± 2/4/8 kV air
± 2/4/6 kV indirect
± 2/4kV direct
± 2/4/8 kV air
Floors should be wood, concrete or ceramic tile. If floors
are covered with synthetic material, the relative
humidity should be at least 30%.
Electrical fast transient/
burst (EFT)
IEC 60601-1-2:2004
o
EN 61000-4-
4:1995 / A1:2001
± 2 kV for power supply
lines
±1 kV for input/output
lines
± 2 kV for power supply
lines
±1 kV for input/output
lines
Mains power should be that of a typical commercial or
hospital environment.
Fast Transient Surge
(FTS)
IEC 60601-1-2:2004
o
EN 61000-4-
5:1995 / A1:2001
± 500V/1 kV
differential mode
± 2 kV common mode
± 500V/1 kV
differential mode
± 2 kV common mode
Mains power should be that of a typical commercial or
hospital environment.
Voltage dips, short
interruptions and voltage
variations on power
supply input lines
IEC 60601-1-2:2004
o
EN 61000-4-
11:1994 / A1:2001
<5% U
t
(>95% dip in U
t
)
for 0.5 cycles
<40% U
t
(>60% dip in
Ut)
for 5 cycles
<70% U
t
(>30% dip in
Ut)
for 25 cycles
<5% U
t
(>95% dip in Ut)
for 5 s
<5% U
t
(>95% dip in U
t
)
for 0.5 cycles
<40% U
t
(>60% dip in
Ut)
for 5 cycles
<70% U
t
(>30% dip in
Ut)
for 25 cycles
<5% U
t
(>95% dip in Ut)
for 5 s
Mains power should be that of a typical commercial or
hospital environment. If the user requires continued
operation during power mains interruptions, it is
recommended that power be supplied from an applicably
rated uninterruptible power supply or a battery.
Power frequency (50/60
Hz) magnetic field
IEC 60601-1-2:2004
o
EN 61000-4-
8:1993 / A1:2001
3 A/m
3 A/m
Power frequency magnetic fields should be at levels
characteristics of a typical location in a typical
commercial or hospital environment.
NOTE:
U
t
is the a.c. mains voltage prior to application of the test level.
Содержание T2100
Страница 1: ...GE T2100 Treadmill Service Manual 2021403 031 Revision A ...
Страница 8: ...vi GE T2100 Treadmill Revision A 2021403 031 ...
Страница 9: ...Revision A GE T2100 Treadmill 1 1 2021403 031 1 Introduction ...
Страница 10: ...1 2 GE T2100 Treadmill Revision A 2021403 031 For your notes ...
Страница 19: ...Revision A GE T2100 Treadmill 2 1 2021403 031 2 Equipment Overview ...
Страница 20: ...2 2 GE T2100 Treadmill Revision A 2021403 031 For your notes ...
Страница 26: ...2 8 GE T2100 Treadmill Revision A 2021403 031 Equipment Overview General Description For your notes ...
Страница 27: ...Revision A GE T2100 Treadmill 3 1 2021403 031 3 Installation ...
Страница 28: ...3 2 GE T2100 Treadmill Revision A 2021403 031 For your notes ...
Страница 44: ...3 18 GE T2100 Treadmill Revision A 2021403 031 Installation Functional Checkout For your notes ...
Страница 45: ...Revision A GE T2100 Treadmill 4 1 2021403 031 4 Maintenance ...
Страница 46: ...4 2 GE T2100 Treadmill Revision A 2021403 031 For your notes ...
Страница 51: ...Revision A GE T2100 Treadmill 5 1 2021403 031 5 Theory of Operation ...
Страница 52: ...5 2 GE T2100 Treadmill Revision A 2021403 031 For your notes ...
Страница 72: ...5 22 GE T2100 Treadmill Revision A 2021403 031 Theory of Operation Electrical Requirements For your notes ...
Страница 73: ...Revision A GE T2100 Treadmill 6 1 2021403 031 6 Troubleshooting ...
Страница 74: ...6 2 GE T2100 Treadmill Revision A 2021403 031 For your notes ...
Страница 102: ...6 30 GE T2100 Treadmill Revision A 2021403 031 Troubleshooting Power Supply Voltage Checks and Fuses For your notes ...
Страница 103: ...Revision A GE T2100 Treadmill 7 1 2021403 031 7 Assembly Disassembly ...
Страница 104: ...7 2 GE T2100 Treadmill Revision A 2021403 031 For your notes ...
Страница 132: ...7 30 GE T2100 Treadmill Revision A 2021403 031 Assembly Disassembly PCBs and Power Supply Replacement For your notes ...
Страница 133: ...Revision A GE T2100 Treadmill 8 1 2021403 031 8 Ordering Parts ...
Страница 134: ...8 2 GE T2100 Treadmill Revision A 2021403 031 For your notes ...
Страница 137: ...Revision A GE T2100 Treadmill A 1 2021403 031 A Appendix A Technical Specifications ...
Страница 138: ...A 2 GE T2100 Treadmill Revision A 2021403 031 For your notes ...
Страница 142: ...A 6 GE T2100 Treadmill Revision A 2021403 031 Appendix A Technical Specifications Technical Specifications ...
Страница 143: ...Revision A GE T2100 Treadmill B 1 2021403 031 B Appendix B Electromagnetic Compatibility EMC ...
Страница 144: ...B 2 GE T2100 Treadmill Revision A 2021403 031 For your notes ...