3 G
10 G
30 G
50 G
100 G
5 G
200 G
Date
Rev
/24
EN-MRI-TYP-SIGNA_EXPLORER-WEB.DWG
A
03/Aug/2018
SIGNA EXPLORER /SIGNA CREATOR
Typical
A3 - Section Views
05
SIDE VIEW WITH MAGNETIC FIELD
Note: Define RF shield's inset according to
provisions made by the RF Shield vendor.
Warning! 5 Gauss line outside
the Magnet room limits.
MAGNET ROOM
CONTROL ROOM
FRONT VIEW WITH MAGNETIC FIELD
6
4
3
2
1
1
3
14
MAGNET ROOM
Note: Define RF shield's inset according to
provisions made by the RF Shield vendor.
Warning! 5 Gauss line outside
the Magnet room limits.
+ 8'-9"
+ TBD
+ TBD
+ 8'-9"
+ TBD
+ TBD
3'-6"
3'-6"
8'-9"
8'-9"