POWER REQUIREMENTS

INTRODUCTION

The system requires two independent power inputs:

main power supply
Magnet monitor power

SPECIFICATIONS OF MAIN POWER INPUT

Power input must be separated from any others which may generate transients (elevators, air conditioning,
radiology rooms equipped with high speed film changers...).

Total harmonic distortion less than 2.5%.

Phase imbalance must not exceed 2%.

SPECIFICATIONS OF MAGNET MONITOR POWER

Power Supply (3
G)

Frequency 50Hz/60Hz

480

415

400

380

208

200

Total Current

60.1

66.8

69.3

72.9

N/A

N/A

Maximum input
power (5 sec max)

59.15 kVA

Stand-by power

13.4 kVA at 0.9 lagging Power Factor including 4.4 KVA for PDU

and 9KVA (continuous operation) for Shield/Cryo Cooler Cabinet.

Critical Power Requirements is different per each configuration.

POWER INPUT

EMERGENCY LIFE-SAFETY POWER SINGLE PHASE + N +G

POWER DEMAND

2.0 A

VOLTAGE

110/220V

FREQUENCY

50 / 60 Hz

CABLES

Power and cable installation must comply with the distribution diagram.
Size of the Main power input cable is determined by the customer, taking its length and admissible voltage
drops  into consideration.

All cables must be isolated and flexible, cable color codes must comply with standards for electrical
installation.
The cables from signaling and remote control (Y,Emergency Off Buttons,L...) will go to the Distribution

Panel/Box with a pigtail length of 1.5m [60in], and will be connected during installation.
Each conductor will be identified and isolated (screw connector).

GROUND SYSTEM

The equipotential link will be by means of an equipotential bar.
The grounding point of Distribution Panel/Box is directly connected to the building's ground by an isolated
copper cable.
The impedance of the earth bar should be less than or equal to 2 ohms

MAIN DISCONNECT PANEL (MDP)

NEUTRAL

GROUND

NEUTRAL MUST

BE TERMINATED
INSIDE THE MDP

NEUTRAL
BUSS BAR

G/PE
BUSS

BAR

CB#1

3-Black
1-White
1-Green

FACILITY

INPUT

POWER

(480-V,

3 Phase)

Cable SUPPLIED BY CUSTOMER

Equipment SUPPLIED BY GE

Equipment SUPPLIED BY
CUSTOMER

Cable SUPPLIED BY GE

POWER DISTRIBUTION

NOTES:

RUNS 296 AND 297, & POWER CORDS FOR SHIELD/CRYO COOLER
COMPRESSOR CABINET, MRCC, & MAGNET MONITOR EQUIPMENT (MAGNET
MONITOR, UPS INPUT & OUTPUT, MODEM, OPTIONAL MULTIPLEXER) ARE
GE SUPPLIED CABLES. ALL OTHER WIRING IS CUSTOMER SUPPLIED.

TWO REMOTE EMERGENCY "OFF" BUTTONS ARE SUPPLIED WITH GE MDP
OPTION, EMERGENCY OFF BUTTONS ARE CUSTOMER SUPPLIED IF GE MDP
OPTION NOT USED.

CIRCUIT BREAKERS ARE PROVIDED FOR PDU, CRYO COOLER COMPRESSOR
CHILLER, GRADIENT CHILLER, SHIELD/CRYO COOLER COMPRESSOR CABINET,
MAGNET MONITOR EQUIPMENT CIRCUITS.

ALL BRANCH CIRCUITS DROP OUT ON LOSS OF POWER. COLDHEAD MRCC,
GRADIENT MRCC, SHIELD/CRYO COOLER COMPRESSOR CABINET, & MAGNET

MONITOR EQUIPMENT AUTOMATICALLY RESTART AFTER 3 SEC TIME DELAY
UPON RESTORATION OF POWER. EMERGENCY OFF LOCKS OUT ALL
CONTACTORS.

IF 3 PHASE WYE WITH NEUTRAL AND GROUND (5 WIRE SYSTEM) INPUT USED

THEN NEUTRAL MUST BE TERMINATED INSIDE THE MAIN DISCONNECT

PANEL AND NOT BROUGHT TO THE POWER CABINET.

SUPERVISORY CIRCUIT FOR HVAC INTERLOCK CONTACTS OPEN ON LOSS OF
DC POWER OR EMERGENCY OFF POSITION.

EQUIPMENT ROOM

"EMERGENCY OFF"

BUTTON

MAGNET ROOM

"EMERGENCY
OFF" BUTTON

PENETRATION

PANEL (PP1)

ARM = AUTO

RESTART MODULE

SHIELD/CRYO COOLER

COMPRESSOR CABINET

(MS5)

GND

FACILITY POWER (24
HOURS/DAY, 7 DAYS/WEEK)
RECOMMENDED TO MAXIMIZE
PROPER UNINTERRUPTED
MAGNET OPERATION FOR
SHIELD/CRYO COOLER
COMPRESSOR CABINET & CRYO
COMPRESSOR CHILLER (MRCC) IF
SUPPLING WATER COOLING FOR
SHIELD/CRYO COOLER
COMPRESSOR CABINET

GROUND STUD OF
SYSTEM CABINET

TO TERMINAL STRIP IN PDU

MODULE LOCATED

BOTTOM OF SYSTEMS

CABINET

POWER DISTRIBUTION UNIT

(PD1)

SYSTEM CABINET

SPLICE

ARM

ARM

CB #3

CB #5

CB #2

EMER-GENCY
OFF MODULE

SUPERVISORY

CKT FOR HVAC

INTERLOCK
CONTACTS

AUTO RESTART

CONTROL

CIRCUIT

120
VAC

FUSES

ARM

CB #3

A

B

C

{

NC

NC

RF SHIELD

RUN 296

RUN 297

BLK

BLK

RED

RED

CB AMPS

1

150

2

100

3

20

4

20

5

30

6

15

7

35

Date

Rev

/24

EN-MRI-TYP-SIGNA_EXPLORER-WEB.DWG

A

03/Aug/2018

SIGNA EXPLORER /SIGNA CREATOR

Typical

E5 - Power Requirements

23

FEEDER TABLE

MIN. feeder wire size,

AWG or MCM (sq. M)/VAC

Minimum feeder wire length - ft (m)

100 (30.5)

150 (46)

200 (61)

250 (76)

300 (92)

350 (107)

400 (122)

450 (137)

480 VAC

1/0 (70)

1/0 (70)

1/0 (70)

1/0 (70)

1/0 (70)

1/0 (70)

1/0 (70)

1/0 (70)

Ground req'd

6 (17)

6 (17)

6 (17)

6 (17)

6 (17)

6 (17)

6 (17)

6 (17)

GENERAL NOTES

In all cases qualified personnel must verify that the feeder (at the point of take-off) and the run to the MR system meet all the requirements

stated in the PIM

For a single unit installation, the minimum transformer size is 112.5KVa.  Regulated transformer is not required unless voltage changes

/- 10% over a period of 1 hour or longer

Grounding conductor will run from the equipment back to  the power source/main grounding point and always travel in the same conduit

with the feeders

Содержание SIGNA CREATOR

Страница 1: ...omplete set of final issue drawing GE cannot accept responsibility for any damage due to the partial use of GE final issue drawings however caused All dimensions are in millimeters unless otherwise specified Do not scale from printed pdf files GE accepts no responsibility or liability for defective work due to scaling from these drawings 24 EN MRI TYP SIGNA_EXPLORER WEB DWG PMM 5538857 1EN 1 4 1 0...

Страница 2: ...plicable Power for MR compressor Chiller is available Ensure cryogen venting system is available for magnet connection Exhaust fan system is installed and operational per GE requirements Status of work General comments System can be delivered PMI signature Site ready for installation FS signature optional GLOBAL SITE READINESS CHECKLIST DI DOC1809666 Rev 5 GENERAL SPECIFICATIONS GE is not responsi...

Страница 3: ...ined to the magnet room If not possible a barrier is recommended to prevent entry to the 5g field area 2 The spaces around above and below the magnet must be reviewed for effects of the 5g 3g 1g and 5g fields Refer to the proximity limit chart in the MR pre installation manual referenced on the cover sheet 3 For moving metal the restriction lines typically extend outside of the MR space Please con...

Страница 4: ... contingent on a 72 in corridor width C 26 Minimum opening for is 43 in w x 82 in h contingent on a 96 in corridor width C 27 Counter top for equipment provide grommeted openings as required to route cables C 28 Base cabinet for storage of surface coils patient positioning pads phantoms etc C 29 Magnet access 9 0 x10 0 C 30 Shelf LEGEND A GE Supplied C Customer contractor supplied and installed B ...

Страница 5: ...Define RF shield s inset according to provisions made by the RF Shield vendor Warning 5 Gauss line outside the Magnet room limits MAGNET ROOM CONTROL ROOM FRONT VIEW WITH MAGNETIC FIELD 6 4 3 2 1 1 3 14 MAGNET ROOM Note Define RF shield s inset according to provisions made by the RF Shield vendor Warning 5 Gauss line outside the Magnet room limits 8 9 TBD TBD 8 9 TBD TBD 3 6 3 6 8 9 8 9 ...

Страница 6: ...gency generators Commercial laundry equipment Food preparation area Water cooling equipment HVAC equipment Major mechanical equipment room Credit cards watches and clocks Air conditioning equipment Fuel storage tanks Motors greater than 5 horsepower 50 gauss 5mT Metal detector for screening LCD panels Telephones No Limit Digital Detectors SOUND PRESSURE SPECTRAL DISTRIBUTION 20 25 31 5 40 50 63 80...

Страница 7: ...inet cover is adjustable between 100 200mm 4 8 in D PP cover is adjustable between 300 400mm 11 8 15 7 in SYSTEM CABINET Top View PENETRATION PANEL Top View D D PENETRATION PANEL Front View Cut out center line RF Shield Finished wall cut out Finished wall RF Shield 125 10 0 5 in 0 in 0 in 1000 0 30 39 in 0 in 1 in 50 10 0 2 in 0 in 0 in 625 10 0 25 in 0 in 0 in 550 10 0 22 in 0 in 0 in 25 10 0 1 i...

Страница 8: ...20 4 in 401 15 79 in 750 29 52 in 700 27 56 in 377 14 84 in 330 12 98 in 208 8 19 in GLOBAL OPERATOR CABINET GOC Center of Gravity SCALE 1 10 TOP VIEW FRONT VIEW SIDE VIEW 300 11 8 in PENETRATION PANEL SIDE VIEW FRONT VIEW SIDE VIEW FRONT VIEW Cover Cover SCALE 1 10 250 9 8 in 300 11 8 in 300 11 8 in 244 9 6 in SYSTEM CABINET Center of gravity SIDE VIEW FRONT VIEW TOP VIEW Air flow 482 19 in 1100 ...

Страница 9: ...gravity GAS SUPPLY RETURN WATER OUT IN 232 9 13 in 450 17 72 in 103 4 06 in 276 10 87 in 153 6 02 in 450 17 72 in 800 31 5 in 153 6 02 in 153 6 02 in BLOWER BOX A Air in take clearance B Duct trim cover TOP VIEW FRONT VIEW SIDE VIEW Air flow A B B 381 15 in 280 11 in 381 15 in 102 4 in TOP VIEW SIDE VIEW FRONT VIEW SCALE 1 25 LOW HEIGHT PATIENT TRANSPORT TABLE FIX 2134 84 in 127 5 in Date Rev 24 E...

Страница 10: ... ROUTING Floor must be able to withstand a moving load of 5320 daN Minimum Opening Height 2 5m 98 5 in width 2 5m 98 5 in Maximum slope 30 degree STRAIGHT PATH Rigging wheels required PATH WITH 90 DEGREE TURN INSTALLATION AND DELIVERY ACCEPTANCE A survey of the site established by the customer and GE Healthcare will make the decision for the delivery time This survey of the site a form is made ava...

Страница 11: ...r safety all removable items within the magnet room such as faucet handles drain covers switch box cover plates light fixture components mounting screws etc Must be non magnetic If you have a specific question about material bring it to the attention of your GE project manager of installations Floor levelness refer to MAGNET ROOM FLOOR SPECIFICATIONS DETAIL this floor levelness requirement is impo...

Страница 12: ...ing 4 Structural wall backing for Magnet Rundown Unit 5 Structural wall backing for Main Disconnect Panel 6 Structural wall backing for Magnet Monitor 7 Floor levelness area for Systems Cabinet Date Rev 24 EN MRI TYP SIGNA_EXPLORER WEB DWG A 03 Aug 2018 SIGNA EXPLORER SIGNA CREATOR Typical S2 Structural Layout 12 MAGNET ROOM EQUIPMENT ROOM CONTROL ROOM 5 4 3 1 1 1 1 7 2 6 0 9 4 11 6 6 4 8 10 6 6 8...

Страница 13: ...nded to the structural elements and not to common screed for magnet anchoring table dock anchor hole VibroAcoustic Pad weight 8 kg 17 lbs each Seismic anchor holes 1346 53 in 673 26 5 in 250 9 8 in 175 6 9 in 1354 53 3 in 76 3 in PATIENT TABLE DOCK ANCHOR MOUNTING REQUIREMENTS The RF Shield vendor must design and install the dock anchor bolt The dock anchor hole must be drilled after the Magnet is...

Страница 14: ...e supplied and installed by the customer or his contractors For complete design and requirements specifications and guidelines refer to the pre installation manual system cooling cryogen venting waveguides and exhaust venting An emergency water cooling back up supply is recommended for continuous cryogen compressor operation if using an open loop back up design ensure a drain is provided please re...

Страница 15: ...Exhaust Intake Vent MAGNET ROOM CONTROL ROOM RF Filter ON OFF Manual fan switches in parallel Ductwork Ductwork MAGNET ROOM VENTING REQUIREMENTS HVAC VENT REQUIREMENTS HVAC vendor must comply with Magnet room temperature and humidity specifications and RF shielding specifications RF Shield vendor must install open pipe or honeycomb HVAC waveguides All serviceable parts in the Magnet room e g diffu...

Страница 16: ...low 30 l min 8 gpm Temperature rise at maximum flow Temperature rise at minimum flow Maximum inlet pressure of Chiller Cryo Compressor 690 kPa 100 psi Minimum inlet pressure of Chiller Cryo Compressor 200 kPa 29 psi Minimum continuous heat load 7 5 kW pH level 6 5 to 8 2 Total hardness Less than 200 ppm of calcium carbonate Suspended matter Less than 10 mg per liter and less than 150 micron partic...

Страница 17: ... PIPE DETAIL MAGNET VENT ADAPTER Do not remove or modify the vent adaptor bolted to the magnet VENT PIPE 610 mm 24 in may be cut to a minimum of 100 mm 4 in VENTGLAS CONNECTOR WAVEGUIDE Additional pipe tubing 0 5 in mesh DIELECTRIC BREAK CRYOGENIC VENTING EXTERIOR GROUND VENT CAP customer supplied OUTSIDE WALL ROOF VENT CAP customer supplied NOT TO SCALE EXHAUST AREA 6 1m x 4 6m 20ft x 15ft LxW EX...

Страница 18: ...7 10 63 4 52 31 17 8 22 56 68 2 26 15 58 4 11 28 34 10 in 250mm 0 20 0 6 1 0 03 0 68 0 55 3 79 0 82 5 65 0 27 1 86 0 04 2 83 20 40 6 1 12 2 0 07 1 58 0 82 5 65 1 51 10 41 0 41 2 83 0 75 5 17 40 60 12 2 18 3 0 10 2 26 1 23 8 48 2 19 15 10 0 62 4 27 1 10 7 58 60 80 18 3 24 4 0 12 2 71 1 51 10 41 2 74 18 89 0 75 5 17 1 37 9 45 80 100 24 4 30 5 0 16 3 62 1 92 13 24 3 43 23 65 0 96 6 62 1 71 11 79 12 i...

Страница 19: ...un direct as possible otherwise may result in the need for greater than standard cable lengths refer to the interconnection diagram for maximum usable lengths point to point 8 Conduit turns to have large sweeping bends with minimum radius in accordance with national and local electrical codes 9 A special grounding system is required in all procedure rooms by some national and local codes It is rec...

Страница 20: ...lephone lines network connection Duplex hospital grade dedicated outlet 120 v emergency single phase power 15a Duplex hospital grade dedicated outlet 120 v single phase outlet routed through RF filter Additional Conduit Runs Contractor Supplied and Installed From To Qty Size in Size mm Main Disconnect Panel Facility power 1 as Req d Main Disconnect Panel Power Distribution Unit inside Systems Cabi...

Страница 21: ...YP SIGNA_EXPLORER WEB DWG A 03 Aug 2018 SIGNA EXPLORER SIGNA CREATOR Typical E3 Electrical Elevations 21 MAGNET ROOM EQUIPMENT ROOM 8 9 TBD TBD 5 4 7 9 9 8 CONTROL ROOM 5 4 3 2 TBD TBD TBD TBD TBD TBD 7 4 9 0 7 6 4 5 0 11 7 6 4 ...

Страница 22: ...N THE FLOOR NOT TO SCALE Removable coverplate DUCT ON THE WALL Date Rev 24 EN MRI TYP SIGNA_EXPLORER WEB DWG A 03 Aug 2018 SIGNA EXPLORER SIGNA CREATOR Typical E4 Electrical Details 22 CABLE MANAGEMENT Removable coverplate WALL DUCT NOT TO SCALE ...

Страница 23: ...TE EMERGENCY OFF BUTTONS ARE SUPPLIED WITH GE MDP OPTION EMERGENCY OFF BUTTONS ARE CUSTOMER SUPPLIED IF GE MDP OPTION NOT USED CIRCUIT BREAKERS ARE PROVIDED FOR PDU CRYO COOLER COMPRESSOR CHILLER GRADIENT CHILLER SHIELD CRYO COOLER COMPRESSOR CABINET MAGNET MONITOR EQUIPMENT CIRCUITS ALL BRANCH CIRCUITS DROP OUT ON LOSS OF POWER COLDHEAD MRCC GRADIENT MRCC SHIELD CRYO COOLER COMPRESSOR CABINET MAG...

Страница 24: ...15 75m 13 94 51 67 ft 17 98 m 59 ft 9 56 m 31 ft 12 5 m 41 ft 12 8m 42 ft 17 4m 57 ft 2 55 16 13m 8 35 52 9 ft 12 2m 40 ft 21 00 m 68 ft 4 3 11 3m 14 1 37 1 ft GOC MAGNET ROOM CONTROL ROOM EQUIPMENT ROOM MRU SC MG6 WC1 CRY MON WC2 PP RF CGS MAG PED PT1 System power panel MDP or PDB SEO2 SEO1 DS Customer supplied 24 m 78 7 ft Date Rev 24 EN MRI TYP SIGNA_EXPLORER WEB DWG A 03 Aug 2018 SIGNA EXPLORE...

Отзывы: