GE Healthcare
Senographe SecondLook Digital CAD System
Revision 1
Operator Manual 5189820-5-C-1EN
CAD.fm
Page no. 13
CAD Image workflow
Image from storage or
archiving device (PACS,
RADSTORE, etc.)
Step 1. The raw image is pushed to the SLDU for analysis.
The image (raw and/or processed) must also be pushed to the review workstation before CAD results can be
viewed.
The user first views the processed image on the review workstation without CAD markers.
Original (raw) image created by a
licensed AWS
AWS
Raw or processed
image
review monitors
Raw or processed
image
CAD overlay
Image displayed
with overlay
review monitors
Step 2. When the SLDU has performed its analysis on each image, it
sends an overlay to the review workstation, showing the location and
type of any regions of interest that were found.
Step 3. While viewing the image, the user presses the CAD button or
the U2 button on the dedicated keypad (if configured) to display the
CAD overlay.
SLDU
AWS
CAD overlay
Raw and/or processed images
SLDU
Raw and/or processed images