Panda iRes Warmer
M1110737 006
© 2007 General Electric Company, all rights reserved
C-2
Manufacturer’s guidance and declaration regarding electromagnetic
immunity
The Giraffe warmer is intended for use in the electronic environment specified below. The user of the
Giraffe warmer should ensure that it is used in such an environment.
Electromagnetic immunity
Immunity Test
IEC 60601 Test
Level
Compliance level
Electromagnetic environment
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
+/- 6kV contact
+/- 8kV air
+/- 6kV contact
+/- 8kV air
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material
the relative humidity should be
at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
+/- 2kV for power
supply line.
+/- 1kV for input/
output line.
+/- 2kV for power
supply line.
+/- 1kV for input/
output line.
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-11
+/- 1kV differential
Mode.
+/- 2kV common
mode line.
+/- 1kV differential
Mode.
+/- 2kV common
mode line.
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage
dips, short
interruptions
and voltage
variations on
power supply
input lines IEC
61000-4-11
<5 % Ut(>95 % dip
in Ut) for 0.5 cycle
40 %Ut (60 % dip
in Ut) for 5 cycles
70 % Ut(30 % dip
in Ut) for 25 cycles
<5% Ut(>95% dip
in Ut) for 5 sec.
<5 % Ut(>95 % dip in
Ut) for 0.5 cycle
40 %Ut(60 % dip in Ut)
for 5 cycles
70 % Ut (30 % dip in
Ut) for 25 cycles
<5% Ut (>95% dip in
Ut) for 5 sec.
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the Giraffe warmer requires
continued operation during
power mains interruptions, it is
recommended that the Giraffe
warmer be powered from an
uninterruptible power supply or
a battery.
Power frequency
(50/60 Hz)
magnetic field
environment
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital.
NOTE:
Ut is the main voltage before application of the test level.