Alarm Table
Alarm
Activation Criteria
Alarm Silence
Audio
Signal*
Mode
Check Baby
Radiant % power @
100% for > 12 min.
Radiant heater %
above “check baby
alarm disabled” limit
for > 12 min.
12 min.
12 min.
2
2
Baby Mode
Manual Mode
Check Baby -
Heat off
Radiant % power @
100% for >15 min. and
the “Check Baby” alarm
has gone unanswered
for three minutes.
Radiant heater %
above “check baby
alarm disabled” limit
for > 15 min. and the
“Check Baby” alarm
has gone unanswered
for three minutes.
15 min.
15 min.
1
1
Baby Mode
Manual Mode
Baby Cold -
Check Temp.
Probe
<2.0° C from Set Temp.
<1.0° C (<0.5° C) from
Set Temp.
5 min.
12 min.
1
2
Baby Mode
Baby Mode
Baby Hot -
Check Temp.
Probe
>2.0° C from Set Temp.
> 1.0° C (>0.5° C) from
Set Temp.
5 min.
12 min.
1
2
Baby Mode
Baby Mode
Confirm Probe
Jack Connection
No longer getting a
reading from the tem
-
perature probe.
2 min.
2 to 1 (after
30 secs.)
Baby Mode or Manual Mode
when Manual Temperature
Alarm set.
Temp. Probe
Failure
Two thermistors in a
probe differ by 0.5° C or
more.
2 min.
2 to 1 (after
30 secs.)
Baby Mode or Manual Mode
when Manual Temperature
Alarm set.
* 1= high priority audio signal
2= low priority audio signal