PanaFlow™ XMT1000 User’s Manual
9
Chapter 1. Installation
1.5.1 Wiring the Analog Outputs
The standard configuration of the XMT1000 flow transmitter includes one isolated 4-20 mA analog output.
Connections to this output may be made with standard twisted-pair wiring, but the current loop impedance for this
circuit must not exceed 600 ohms. Two additional analog outputs are available as an option.
To wire the analog outputs, complete the following steps:
1.
Disconnect the main power to the unit and remove the wiring cover.
2.
Install the required cable gland in the chosen conduit hole on the side of the electronics enclosure.
3.
for the location of the terminal block and wire the analog output as shown. Secure
the cable gland.
IMPORTANT:
Analog output A is an active signal. Do not supply power to this circuit, as the circuit is powered by the
flow meter.
4.
If wiring of the unit has been completed, reinstall the wiring cover on the enclosure and tighten the set screw.
Note:
Prior to use, the analog output must be set up and calibrated. Proceed to the next section to continue the initial
wiring of the unit.
Note:
See Appendix A,
Specifications,
for the load and voltage requirements.
Attention European Customers!
To meet CE Marking requirements, all cables must be installed as
“CE Marking Compliance” on page 1
WARNING!
Make sure all covers, with their o-ring seals, are installed and the set screws tightened
before applying power in a hazardous environment.
Содержание PanaFlow XMT1000
Страница 1: ...GE Oil Gas Flow 910 313 Rev A June 2017 PanaFlow XMT1000 User s Manual ...
Страница 2: ......
Страница 4: ...ii no content intended for this page ...
Страница 8: ...Contents vi PanaFlow XMT1000 User s Manual no content intended for this page ...
Страница 12: ...Preface x PanaFlow XMT1000 User s Manual no content intended for this page ...
Страница 42: ...Chapter 3 Error Codes and Troubleshooting 30 PanaFlow XMT1000 User s Manual no content intended for this page ...
Страница 46: ...34 PanaFlow XMT1000 User s Manual no content intended for this page ...
Страница 68: ...Appendix C HART Communication 56 PanaFlow XMT1000 User s Manual no content intended for this page ...
Страница 95: ......