Z Series RO System by GE Osmonics
5.9
Product Flow Too High
1.
Rejection Satisfactory
A.
Waste Pressure Too High
a. Adjust pressure regulator.
b. Waste flow too low.
B.
Feed Water Temperature Too High
a. Adjust temperature blending valve.
b. Correct plumbing error that allows the cold water to be heated.
2.
Rejection Unsatisfactory
A.
ROMA Internal Seals Damaged, Dirty, Or Dislodged
a. Disassemble ROMA, clean and inspect “O” rings and seals.
B.
Membrane Hydrolyzed (Exposed To Oxidant)
a. Verify condition of pretreatment equipment (GAC tank).
b. Replace membrane.
C.
Waste Flow Too Low
a. Adjust
flow.
5.10 Poor Apparent Water Quality
1.
Waste Flow Too Low
a. Open waste flow control valve.
2.
Waste Pressure Too Low
a. Adjust waste pressure regulator.
3.
Product Flow Too High
a. See section 5.9.
4.
Membrane Scaled Or Fouled
a. Verify condition of pretreatment equipment.
b. Reference the membrane cleaning procedure to remove scale
5.
Erroneous Meter Readout
a. Verify water quality with independent meter.
b. Feed or product water probe wire disconnected or loose.
c. Replace feed and/or product water probe.
d. Replace monitor circuit board.
e. Replace product water probe.
f. Check or replace TDS temperature probe (10,000 Ohms @ 25 C).
6.
Membrane Failure
a. Replace
membrane.
7.
Change In Feed Water TDS
a. Verify water quality with independent meter.
1238339b – 16Oct03
5 - 9
Troubleshooting
Содержание Osmonics Z-10000
Страница 2: ......
Страница 6: ...This page intentionally left blank 1238339b D D 16 Oct 2003 ...
Страница 27: ...Z Series RO System by GE Osmonics NOTES 1238339b 16Oct03 1 21 General Information ...
Страница 28: ...Z Series RO System by GE Osmonics This page intentionally left blank 1238339b 16Oct03 1 22 General Information ...
Страница 37: ...Z Series RO System by GE Osmonics NOTES 1238339b 16Oct03 2 9 Installation Guidelines and Instructions ...
Страница 40: ...Z Series RO System by GE Osmonics This page intentionally left blank 1238339b 16Oct03 3 2 System Operation ...
Страница 43: ...Z Series RO System by GE Osmonics NOTES 1238339b 16Oct03 3 5 System Operation ...
Страница 44: ...Z Series RO System by GE Osmonics This page intentionally left blank 1238339b 16Oct03 3 6 System Operation ...
Страница 46: ...Z Series RO System by GE Osmonics This page intentionally left blank 1238339b 16Oct03 4 2 Routine Maintenance ...
Страница 72: ...Z Series RO System by GE Osmonics NOTES 1238339b 16Oct03 4 28 Routine Maintenance ...
Страница 74: ...Z Series RO System by GE Osmonics This page intentionally left blank 1238339b 16Oct03 5 2 Troubleshooting ...
Страница 76: ...Z Series RO System by GE Osmonics This page intentionally left blank 1238339b 16Oct03 5 4 Troubleshooting ...
Страница 84: ...Z Series RO System by GE Osmonics NOTES 1238339b 16Oct03 5 12 Troubleshooting ...
Страница 85: ...Z Series RO System by GE Osmonics Z Series RO System CHAPTER SIX DRAWINGS 1238339b 16Oct03 6 1 Drawings ...
Страница 86: ...Z Series RO System by GE Osmonics This page intentionally left blank 1238339b 16Oct03 6 2 Drawings ...
Страница 96: ...Z Series RO System by GE Osmonics 1238339b 16Oct03 6 12 Drawings Z Series RO Indirect Feed Flow Diagram ...
Страница 97: ...Z Series RO System by GE Osmonics 1238339b 16Oct03 6 13 Drawings Z Series RO Direct Feed Flow Diagram ...
Страница 101: ...Z Series RO System by GE Osmonics 1238339b 16Oct03 6 17 Drawings I O OFF ON Z Series RO Typical Controller ...
Страница 106: ...Z Series RO System by GE Osmonics Z Series RO Disinfect Alarm Monitor Wiring Connections 1238339b 16Oct03 6 22 Drawings ...
Страница 107: ...Z Series RO System by GE Osmonics 1238339b 16Oct03 6 23 Drawings Z Series RO Fluid Panel ...
Страница 110: ...Z Series RO System by GE Osmonics 1238339b 16Oct03 6 26 Drawings Z Series RO Waste Flow Control Valve Assembly ...
Страница 114: ...Z Series RO System by GE Osmonics Over Temperature Probe Probe Installation 1238339b 16Oct03 6 30 Drawings ...
Страница 119: ...Z Series RO System by GE Osmonics NOTES 1238339b 16Oct03 6 35 Drawings ...
Страница 120: ...Z Series RO System by GE Osmonics This page intentionally left blank 1238339b 16Oct03 6 36 Drawings ...
Страница 122: ...Z Series RO System by GE Osmonics This page intentionally left blank 1238339b 16Oct03 7 2 Spare Parts List ...
Страница 126: ...Z Series RO System by GE Osmonics NOTES 1238339b 16Oct03 7 6 Spare Parts List ...
Страница 127: ......