GE C
OMPANY
D
IRECTION
5472001-1EN, R
EVISION
6
O
PTIMA
CT680 S
ERIES
AND
O
PTIMA
CT670 I
NSTALLATION
M
ANUAL
Page 54
Section 9.0 - Table Installation (GT1700V)
Section 9.0 Table Installation (GT1700V)
Note:
For Lite Table Installation, refer to
.
9.1
Time and Personnel
9.2
Tools and Test Equipment
•
Standard Install Tool Kit
•
3/4", 1-1/4", 1-1/2" and 1-5/8" sockets
•
8mm, 10mm, and 14mm hex socket bits
•
Laser Alignment kit
•
Johnson Professional 6" level
•
Johnson Professional 4’ level
•
Johnson Professional 2’ level
9.3
Procedures
9.3.1
Draw Table Reference Lines
1.) Draw a reference line of 673 ± 6 mm (26.5" ± 0.25")position from Gantry Base on the floor as
. This line should be parallel to the gantry. In a later section, you will
move the table against the 673 mm (26.5") mark.
Figure 1-25 Draw Reference Line
Required Persons
Preliminary Reqs
Procedure
Finalization
2 (FE or mechanical supplier)
1.5 hours labor on-site
Draw this line on the floor.
Gantry Base
673 ± 6 mm (26.5" ± 0.25")