M
Millenium™ RO System by GE Osmonics
14693e – 31 Mar 03
1 - 12
General Information
1.10 Specifications:
1.10.1 Input Water Requirements
Minimum
Maximum
Input
flow
Rate 2.0
GPM
N/A
Temperature
(water)
1.7°C
(35°F)
32°C
(90°F)
NOTE
:
Operating at lower temperatures may make it impossible for the Millenium™ to
produce adequate permeate for the dialysis machine.
pH
(operating)
4.0
11.0
TF
Membranes
pH w/chloramines present (operating)
4.0
8.5
TF
Membranes
pH
(cleaning)
2.0
11.0
TF
Membranes
Pressure,
dynamic
30
PSI
100
PSI
1.10.2 Flow
Rates
Product Flow based on the following conditions:
New
Membranes
Operating
Pressure:
190
PSI
Product
Pressure:
1
PSI
Feed Water Temperature:
77
°
F (25
°
C)
GPM LPM GPD LPD
.34 1.31 500 1892.5
.52 1.97 750 2839.1
Membrane flux may vary ±15%
Waste Flow: .84 GPM (3.2 LPM) (
±
10%)
1.10.3 Electrical
Requirements
Motor
1/3 HP
Voltage 115
VAC
/
230
VAC
Hertz
60
/
50
Amps
8
/
5
Phase
single
Содержание MILLENIUM OSMONICS MILLENIUMTM REVERSE OSMOSIS UNIT
Страница 2: ......
Страница 6: ...14693e D D 31 March 2003 This page intentionally left blank ...
Страница 24: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 1 18 General Information NOTES ...
Страница 36: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 3 5 5 System Operation NOTES ...
Страница 44: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 4 7 Routine Maintenance Tubing connections ...
Страница 67: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 4 30 Routine Maintenance NOTES ...
Страница 69: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 5 2 Troubleshooting This page intentionally left blank ...
Страница 71: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 5 4 Troubleshooting This page intentionally left blank ...
Страница 80: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 5 13 Troubleshooting NOTES ...
Страница 81: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 5 14 Troubleshooting This page intentionally left blank ...
Страница 82: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 6 1 Drawings Millenium RO System CHAPTER SIX DRAWINGS ...
Страница 83: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 6 2 Drawings This page intentionally left blank ...
Страница 84: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 6 3 Drawings Millenium RO Component Layout ...
Страница 85: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 6 4 Drawings Millenium RO Disinfect View ...
Страница 86: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 6 5 Drawings Millenium RO Flow Schematic ...
Страница 96: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 6 15 Drawings Push In Fitting Instructions ...
Страница 97: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 6 16 Drawings This page intentionally left blank ...
Страница 98: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 6 17 Drawings NOTES ...
Страница 99: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 6 18 Drawings This page intentionally left blank ...
Страница 101: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 7 2 Spare Parts List This page intentionally left blank ...
Страница 105: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 7 6 Spare Parts List NOTES ...
Страница 106: ......