will rinse the cell, this time with water, using greater quantities. This is especially useful
after a detergent soak, to be sure of removing all of the detergent.
Click the
Detergent Soak and Rinse (Long)
button and follow the prompts to load ~5%
Contrad70.
The cells will heat to 50°C for half an hour to soak. Once they have cooled, the software
will prompt the user to remove the Contrad70 and insert the cleaning device for rinsing,
and then to connect the pipette connector to its holder. Be sure that the cells have
cooled before removing the Contrad70, as hot liquid may shatter the loading syringe.
This method is used when more vigorous cleaning is required, after a precipitating
sample or if trouble is suspected. It is also a good idea to do this every few weeks, or
every few days for an instrument under heavy use, to prevent problems.
Connect the syringe tubing to the syringe fill port. Click
Syringe Wash (Short)
. The
software will prompt the user to make sure that the tubing is connected.
Click
OK
. The Washing Module will perform a short washing routine (approximately 8
minutes). It will rinse the syringe first with water and then with methanol. It will then pull
vacuum through the syringe for several minutes to dry it.
Connect the syringe tubing to the syringe fill port. Click
Syringe Wash (Long)
. The
software will prompt the user to make sure that the tubing is connected.
Click
OK
. The Washing Module will perform a more thorough washing routine
(approximately 8½ minutes). It will rinse the syringe first with water and then with
methanol. It will then pull vacuum through the syringe for several minutes to dry it.
The
Dry Syringe
function is used when the syringe has been cleaned manually and left
with methanol in it. Attach the syringe connector to the syringe and place the pipette in
the wash position. Click the
Dry Syringe
button.
The washing module will pull vacuum through the syringe for several minutes to dry it.
4.8
Real time plot
This tab displays the current DP data.
See MicroCal iTC
200
Experimental and Data Analysis tutorial Guide.
4.9
Set up
This tab allows for the customizing of data file paths for data storage and setup files. It
is also where the
Extended Data Mode
button can be found. It is useful to have this box
checked when running test experiments. The additional data it provides will be helpful
for troubleshooting. See MicroCal iTC
200
Experimental and Data Analysis tutorial Guide.
Содержание MicroCal iTC200
Страница 1: ...GE Healthcare MicroCal iTC200 System Operating Instructions Original Instructions ...
Страница 2: ......
Страница 16: ...1 Introduction 1 4 Control software 16 MicroCal iTC200 Operating Instructions 28 9639 77 AA ...
Страница 28: ...2 Safety instructions 2 3 Recycling procedures 26 MicroCal iTC200 Operating Instructions 28 9639 77 AA ...
Страница 50: ...4 Operation 4 10 Procedures after a run 48 MicroCal iTC200 Operating Instructions 28 9639 77 AA ...
Страница 63: ......