4 Operation
4.7 Instrument controls
44
MicroCal iTC
200
Operating Instructions
28-9639-77
AA
4.7
Instrument controls
Figure 4-5.
Instrument control screenshow.
The
Instrument Controls
tab contains the controls for direct operation of the
instrument. At the top of the window, the user can
Start
the run, using whatever
parameters are currently present in the
Experimental Design
or
Advanced
Experimental Design
tabs.
Before clicking this button, it is wise to check that all parameters are correct and that a
valid, unique data file name has been entered. The software will double-check with the
user before allowing any files to be overwritten. The
Stop
button, which is available only
during a run, will abort the run immediately.
The
Thermostat Control
section allows for setting of the thermostat temperature,
which will be maintained during the MicroCal iTC
200
thermostatic (idle) state. Pre-
thermostatting the MicroCal iTC
200
and samples at the run temperature will result in
shorter equilibration times. Also, high temperature thermostatting during cell cleaning
can improve the effects of the cleaning. Use the arrow buttons to raise or lower the
temperature or click in the text box and type in a new number. Click
Set Jacket Temp
to
set the temperature.
Содержание MicroCal iTC200
Страница 1: ...GE Healthcare MicroCal iTC200 System Operating Instructions Original Instructions ...
Страница 2: ......
Страница 16: ...1 Introduction 1 4 Control software 16 MicroCal iTC200 Operating Instructions 28 9639 77 AA ...
Страница 28: ...2 Safety instructions 2 3 Recycling procedures 26 MicroCal iTC200 Operating Instructions 28 9639 77 AA ...
Страница 50: ...4 Operation 4 10 Procedures after a run 48 MicroCal iTC200 Operating Instructions 28 9639 77 AA ...
Страница 63: ......