7-4
MAC
1200 resting ECG analysis system
Revision D
2012250-095
Service and Maintenance: Technical Inspections/Preventative Maintenance
Visual Check
Device and accessories should be checked for the following:
Fuse cartridges comply with vendor's specifications.
Labels and inscriptions on the device relating to safety are clearly legible.
The mechanical state of the device permits its further use.
There is no product deterioration (dirtiness, dents, rough edges, etc) which
could cause any reduction in safety.
Test Functions
Recommended Testing Instruments and Accessories
1x
Multi-parameter simulator Lionheart
1x
RS232 interface connector with internal connection between pins 2 and 3
(TXD and RXD)
1x
Patient trunk cable and customer electrode leads, or 1X patient trunk cable, 10-
pin 223 387 01
Test Preparations
In general, the device test functions implemented in MAC
1200 resting ECG
analysis system are used for the tests. See Chapter 3,
for
details.
Connect MAC
1200 resting ECG analysis system up to the mains, the green LED
for standby must glow.
Switch the device on, the self-test runs automatically, no error message should
appear. When the self-test has finished, the device is in the automatic mode, the
yellow LED (indicating disabled operating mode) should still be glowing.
Modifications in the user-programmed configuration may need to be made in order
to carry out the test. If such a change needs to be made to enable testing, make a
printout of all the modified configuration lists and mark the changes made.
NOTE
After completing the test the original user-programmed configuration is to be
retrieved and activated.
Operating and Display Unit Performance Tests
Carry out the
as detailed in Chapter 3,
Carry out the
as detailed in Chapter 3,
Test for Recording Speeds 25 and 50 mm/s
Carry out the
as detailed in Chapter 3,
.
The feed speed deviations should be less than 3%.
Содержание MAC 1200
Страница 4: ...Compliance 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 Compliance For your notes ...
Страница 10: ...vi MAC 1200 resting ECG analysis system Revision D 2012250 095 ...
Страница 11: ...Revision D MAC 1200 resting ECG analysis system 1 1 2012250 095 1 The Basics ...
Страница 12: ...1 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Страница 19: ...Revision D MAC 1200 resting ECG analysis system 2 1 2012250 095 2 Functional Description ...
Страница 20: ...2 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Страница 45: ...Revision D MAC 1200 resting ECG analysis system 3 1 2012250 095 3 System Test Function ...
Страница 46: ...3 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Страница 68: ...3 24 MAC 1200 resting ECG analysis system Revision D 2012250 095 System Test Function Pharma For your notes ...
Страница 69: ...Revision D MAC 1200 resting ECG analysis system 4 1 2012250 095 4 Repair Instructions ...
Страница 70: ...4 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Страница 79: ...Revision D MAC 1200 resting ECG analysis system 5 1 2012250 095 5 Troubleshooting ...
Страница 80: ...5 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Страница 85: ...Revision D MAC 1200 resting ECG analysis system 6 1 2012250 095 6 Adjustment Instructions ...
Страница 86: ...6 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Страница 89: ...Revision D MAC 1200 resting ECG analysis system 7 1 2012250 095 7 Service and Maintenance ...
Страница 90: ...7 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Страница 101: ...Revision D MAC 1200 resting ECG analysis system 8 1 2012250 095 8 Parts List ...
Страница 102: ...8 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Страница 111: ...Revision D MAC 1200 resting ECG analysis system 9 1 2012250 095 9 Specifications ...
Страница 112: ...9 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Страница 122: ...9 12 MAC 1200 resting ECG analysis system Revision D 2012250 095 Specifications Basic Information For your notes ...
Страница 123: ...Revision D MAC 1200 resting ECG analysis system 10 1 2012250 095 10 Drawings ...
Страница 150: ......