Revision D
MAC
1200 resting ECG analysis system
3-5
2012250-095
System Test Function: Display Test
Display Test
After pressing the
1
key, the following menu appears.
Terminate with any key.
Pressing any other key than one of those listed in the menu above leads back to the
initial display menu.
Contrast Control
The contrast setting can always be adjusted as described above (cursor keys).
Table 2. Display Test
Key
Test Performed
Function
1
Test pattern
Pressing this key generates a vertical stripe
pattern. Each subsequent pressing generates the
inverse display of its predecessor.
2
Clear display
All display pixels are inactive.
3
Restore menu
The initial display test menu is displayed.
4
Display illumination
The illumination is switched on or off by pressing
this key.
Contrast control
Shift + Cursor_Down
Shift + Cursor_Up
Содержание MAC 1200
Страница 4: ...Compliance 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 Compliance For your notes ...
Страница 10: ...vi MAC 1200 resting ECG analysis system Revision D 2012250 095 ...
Страница 11: ...Revision D MAC 1200 resting ECG analysis system 1 1 2012250 095 1 The Basics ...
Страница 12: ...1 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Страница 19: ...Revision D MAC 1200 resting ECG analysis system 2 1 2012250 095 2 Functional Description ...
Страница 20: ...2 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Страница 45: ...Revision D MAC 1200 resting ECG analysis system 3 1 2012250 095 3 System Test Function ...
Страница 46: ...3 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Страница 68: ...3 24 MAC 1200 resting ECG analysis system Revision D 2012250 095 System Test Function Pharma For your notes ...
Страница 69: ...Revision D MAC 1200 resting ECG analysis system 4 1 2012250 095 4 Repair Instructions ...
Страница 70: ...4 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Страница 79: ...Revision D MAC 1200 resting ECG analysis system 5 1 2012250 095 5 Troubleshooting ...
Страница 80: ...5 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Страница 85: ...Revision D MAC 1200 resting ECG analysis system 6 1 2012250 095 6 Adjustment Instructions ...
Страница 86: ...6 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Страница 89: ...Revision D MAC 1200 resting ECG analysis system 7 1 2012250 095 7 Service and Maintenance ...
Страница 90: ...7 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Страница 101: ...Revision D MAC 1200 resting ECG analysis system 8 1 2012250 095 8 Parts List ...
Страница 102: ...8 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Страница 111: ...Revision D MAC 1200 resting ECG analysis system 9 1 2012250 095 9 Specifications ...
Страница 112: ...9 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Страница 122: ...9 12 MAC 1200 resting ECG analysis system Revision D 2012250 095 Specifications Basic Information For your notes ...
Страница 123: ...Revision D MAC 1200 resting ECG analysis system 10 1 2012250 095 10 Drawings ...
Страница 150: ......