2-20
MAC
1200 resting ECG analysis system
Revision D
2012250-095
Functional Description: Internal Interfaces
Interface to the Keypad
The interface to the keypad is realized with the connector
KEYB/
. It contains the
signals for the keypad rows and columns, the supply and the control signals for the
status LEDs of the keypad. The foil connecting cable is part of the keypad itself.
Connector denotation:
KEYB/
Type:
Foil connector, 26-pin, zero power insertion, 180°, grid
1.0 mm, Molex
Interface to the Printhead
The control data to the printhead is transferred via the flexible PCB Printhead
Connection, plugged into the connector
TPC_DAT/
on the PCB Control CS_CI.
Connector denotation:
TPC_DAT/
Type:
Male connector 1 x 20-pin, 180°
Printhead Supply
The printhead is supplied through a 6-pin connection cable, plugged to the connector
TPC_PO
on the PCB Control CS_CI.
Connector denotation:
TPC_PO
Type:
Multipoint connector 1 x 6 pin, 180°,
Reverse terminal protection, MODU II
The function of the individual pins is given in the table below.
TPC_PO
Pin Number
Signal
Function
6
COMMON
Supply V22.5 V
5
GNDITPC
Ground
4
COMMON
Supply V22.5 V
3
COMMON
Supply V22.5 V
2
GNDITPC
Ground
1
+22.5V
Auxiliary Voltage
Содержание MAC 1200
Страница 4: ...Compliance 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 Compliance For your notes ...
Страница 10: ...vi MAC 1200 resting ECG analysis system Revision D 2012250 095 ...
Страница 11: ...Revision D MAC 1200 resting ECG analysis system 1 1 2012250 095 1 The Basics ...
Страница 12: ...1 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Страница 19: ...Revision D MAC 1200 resting ECG analysis system 2 1 2012250 095 2 Functional Description ...
Страница 20: ...2 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Страница 45: ...Revision D MAC 1200 resting ECG analysis system 3 1 2012250 095 3 System Test Function ...
Страница 46: ...3 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Страница 68: ...3 24 MAC 1200 resting ECG analysis system Revision D 2012250 095 System Test Function Pharma For your notes ...
Страница 69: ...Revision D MAC 1200 resting ECG analysis system 4 1 2012250 095 4 Repair Instructions ...
Страница 70: ...4 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Страница 79: ...Revision D MAC 1200 resting ECG analysis system 5 1 2012250 095 5 Troubleshooting ...
Страница 80: ...5 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Страница 85: ...Revision D MAC 1200 resting ECG analysis system 6 1 2012250 095 6 Adjustment Instructions ...
Страница 86: ...6 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Страница 89: ...Revision D MAC 1200 resting ECG analysis system 7 1 2012250 095 7 Service and Maintenance ...
Страница 90: ...7 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Страница 101: ...Revision D MAC 1200 resting ECG analysis system 8 1 2012250 095 8 Parts List ...
Страница 102: ...8 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Страница 111: ...Revision D MAC 1200 resting ECG analysis system 9 1 2012250 095 9 Specifications ...
Страница 112: ...9 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Страница 122: ...9 12 MAC 1200 resting ECG analysis system Revision D 2012250 095 Specifications Basic Information For your notes ...
Страница 123: ...Revision D MAC 1200 resting ECG analysis system 10 1 2012250 095 10 Drawings ...
Страница 150: ......