GE
RAFT
LOGIQ P9/P7
D
IRECTION
5604324
, R
EVISION
11
DRAFT (J
ANUARY
24, 2019)
S
ERVICE
M
ANUAL
1-10
Section 1-3 - Safety Considerations
1-3-5
Auxiliary Devices Safety
The LOGIQ P9/P7 may be used with an isolation transformer to provide the required separation from
mains for both, the system and the auxiliary devices.
One AUX main outlet is located at the primary power supply. It is used for connecting the two-fold splitter
whose outlets are led to the shelves intend for auxiliary devices (e.g., printers) and the AUX main outlet
that is accessible on the back of the control console.
The IEC 60601-1-1 standard provides a guideline for safely interconnecting medical devices in systems.
“Equipment connected to the analog or digital interface must comply with the respective IEC/UL
standards (e.g. IEC 60950 / UL 60950 for data processing equipment and IEC 60601-1 / ANSI/AAMI
ES60601-1 for medical equipment).
Everybody who connects additional equipment to the signal input portion or signal output portion
configures a medical system, and is therefore responsible that the system complies with the
requirements of the system standard IEC 60601-1-1.
Special care has to be taken, if the device is connected to computer network (e.g., Ethernet), because
other devices could be connected without any control. There could be a potential difference between
the protective earth and any line of the computer network including the shield.
In this case the only way to operate the system safely is to use an isolated signal link with minimum
4mm creepage distance, 2.5mm air clearance of the isolation device. For computer networks there are
media converters available which convert the electrical to optical signals. Please consider that this
converter has to comply with IEC xxx standards
*
and is battery operated or connected to the isolation
mains output of the LOGIQ P9/P7 ultrasound system.
*
IEC xxx stands for standards such as:
• IEC 60601 for medical devices
• IEC 60950 for information technology equipment etc.
WARNING
WARNING
!! WARNING:
Power Supplies for additional equipment MUST comply with IEC 60601-1.
WARNING
WARNING
!! WARNING:
DO NOT attempt to use different peripherals and accessories (brand and model;
connected via USB ports) other than approved and provided by GE Healthcare!
The ultrasound system is an extremely sensitive and complex medical system.
Any unauthorized peripherals may cause system failure or damage!
NOTICE
!! NOTICE:
The system integrator
(any person connecting the medical device to other devices)
is responsible
that the connections are safe.
If in doubt, consult the technical service department or your local representative.
CAUTION
!! CAUTION:
The leakage current of the entire system including any / all auxiliary equipment must not exceed
the limit values as per IEC 60601-1 / EN 60601-1 respectively other valid national or international
standards. All equipment must comply with UL, CSA and IEC requirements.
Содержание LOGIQ P9
Страница 2: ......
Страница 11: ...GE LOGIQ P9 P7 DIRECTION 5604324 REVISION 11 SERVICE MANUAL ix ZH CN KO ...
Страница 40: ...GE LOGIQ P9 P7 DIRECTION 5604324 REVISION 11 SERVICE MANUAL xxxviii Table of Contents ...
Страница 184: ...GEDRAFT LOGIQ P9 P7 DIRECTION 5604324 REVISION 11 DRAFT JANUARY 24 2019 SERVICE MANUAL Chapter 4 Functional Checks 4 44 ...
Страница 601: ...This page was intentionally left blank ...