GE H
EALTHCARE
D
IRECTION
5245279, R
EVISION
3
LOGIQ™ P6/P6 P
RO
S
ERVICE
M
ANUAL
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8 - 67
8-2-27
P3RLY/PI128II Assy
8-2-27-1
Tools
•
Common philips screwdrivers
8-2-27-2
Preparations
•
Shut down the system and switch off the main breaker.
8-2-27-3
Removal procedure
1.) Remove the side left and right covers. Refer to the
8-2-14 "Right or Left Side Cover" on page 8-47
..
2.) Remove the Front Cover. Refer to the
8-2-15 "Front Cover" on page 8-49
.
3.) Remove the RFC Assy. Refer to the
8-2-32 "RFC Assy" on page 8-77
4.) Unscrew 8 screws (1-8) and remove the metal cover. Refer to the Figure 8-107.
5.) Unscrew 6 screws (1~6, white arrow marked). Refer to the Figure 8-108.
Figure 8-107 P3RLY/PI128II Cover
Figure 8-108 P3RLY/PI128II Assy