10-3-1 How often should care & maintenance tasks be performed? (cont’d)
Abbreviations used in the Customer Care Schedule
:
D = Daily
W = Weekly
M = Monthly
A = Annually
Table 10-2 Customer Care Schedule
Item
Service at Indicated Time
D
W
M
A
Notes
Air Filter Grid
Remove the filter grid and clean the air filter.
•
Or at least bi-weekly
AC Mains Cable
Inspect AC Mains Cable
•
Mobile Unit Check weekly
Cables and Connectors
Check if all cables are fixed well seated at
the correct position and if there is no
mechanical damage visible.
•
also after corrective
maintenance
User Interface
Clean alphanumerical keyboard, Functional
keys, Digital potentiometers, TGC-Shift
potentiometers. (vacuum cleaner, lukewarm
soap water on a soft, damp cloth)
•
Be careful not to get the cloth
too wet so that moisture does
not enter the loudspeakers,
TGC-Slider, or other keys!
LCD Monitor, Touch
Panel and Probe holder
Clean LCD Monitor surface and Probe
holder with a fluid detergent in warm water
on a soft, damp cloth.
•
Be careful not to get the cloth
too wet so that moisture does
not enter the entire system.
Mechanical parts
Clean and inspect the mechanical function
of wheels, casters, brakes and swivel locks
as well as side door, foot rest, front and rear
handle, and monitor holder.
Remove Dust and Coupling gel.
•
Mobile Unit Check Daily
Control Console
movement
Check Translation/Rotation and Height
Adjustment (Elevation)
•
more frequently at Mobile Units
Trackball Check
Check proper operation
(Cursor movement X, Y direction)
•
If failure occurs go to trackball
cleaning.
Trackball Cleaning
Remove trackball ring; open the trackball
housing and take out the trackball to clean it
with soft tissue and screwdriver shaft.
•
Please record it in the systems
setup maintenance report
Disk Drives
(Data Backup)
Test Image filing (Archive) Import and
Export data capability (DVD/CD Drive)
•
•*
* save the image filing data
weekly or at least monthly on
DVD/CD depending on the
number of examinations
Safe Probe Operation
Clean probes and probe cables and check
acoustic lens housing (cracks) and probe
cables. In case of mechanical damage,
don’t use them!
Danger:
Safety risk for
operator and patient.
•*
* or before each use
Probe Air bubbles
To detect air bubbles in filling liquid, shake
the probe carefully and check abnormal
noise.
•
Probe connectors
Remove dust/dirt of all probe connectors.
Clean with vacuum cleaner if dust is visible.
•
Console Leakage
Current Checks
•
Also after corrective
maintenance or as required by
your facilities QA program.
Peripheral Leakage
Current Checks
•
Also after corrective
maintenance or as required by
your facilities QA program.
Содержание LOGIQ E10
Страница 1: ...LOGIQ E10 Basic Service Manual 5750007 1EN Rev 1 Version R1 2018 GENERAL ELECTRIC COMPANY...
Страница 2: ......
Страница 3: ...DIRECTION 5750007 1EN REV 1 LOGIQ E10 BASIC SERVICE MANUAL i Important Precautions TRANSLATION POLICY...
Страница 4: ...DIRECTION 5750007 1EN REVISION REV 1 LOGIQ E10 BASIC SERVICE MANUAL ii...
Страница 5: ...DIRECTION 5750007 1EN REV 1 LOGIQ E10 BASIC SERVICE MANUAL iii...
Страница 6: ...DIRECTION 5750007 1EN REVISION REV 1 LOGIQ E10 BASIC SERVICE MANUAL iv...
Страница 7: ...DIRECTION 5750007 1EN REV 1 LOGIQ E10 BASIC SERVICE MANUAL v...
Страница 8: ...DIRECTION 5750007 1EN REVISION REV 1 LOGIQ E10 BASIC SERVICE MANUAL vi...
Страница 9: ...DIRECTION 5750007 1EN REV 1 LOGIQ E10 BASIC SERVICE MANUAL vii...
Страница 10: ...DIRECTION 5750007 1EN REVISION REV 1 LOGIQ E10 BASIC SERVICE MANUAL viii...
Страница 11: ...DIRECTION 5750007 1EN REV 1 LOGIQ E10 BASIC SERVICE MANUAL ix...
Страница 12: ...DIRECTION 5750007 1EN REVISION REV 1 LOGIQ E10 BASIC SERVICE MANUAL x...
Страница 13: ...DIRECTION 5750007 1EN REV 1 LOGIQ E10 BASIC SERVICE MANUAL xi...
Страница 14: ...DIRECTION 5750007 1EN REVISION REV 1 LOGIQ E10 BASIC SERVICE MANUAL xii...
Страница 18: ...DIRECTION 5750007 1EN REVISION REV 1 LOGIQ E10 BASIC SERVICE MANUAL xvi This page was intentionally left blank...
Страница 34: ...DIRECTION 5750007 1EN REV 1 LOGIQ E10 BASIC SERVICE MANUAL xxxii Hardware Kit 9 49...
Страница 776: ...DIRECTION 5750007 1EN REV 1 LOGIQ E10 BASIC SERVICE MANUAL Index 4...
Страница 777: ......
Страница 778: ...2018 General Electric Company GE Healthcare 9900 Innovation Drive RP 2156 Wauwatosa WI 53226 USA www gehealthcare com...