Section 10-7 - Electrical Safety Tests
10-7-11
Mains on applied part
NOTE:
Mains Applied refers to the sink leakage test where mains (supply) voltage is applied to the part to
determine the amount of current that will pass (or sink) to ground if a patient contacted mains voltage.
Mains on applied part is one of the described leakage current tests applicable for probes (Ref:
IEC60601-1). This is to be performed with the probe disconnected from the Ultrasound system. Apply
mains voltage over the insulation barrier. (Between protective earth on the probe connector, and an
electrical anode in saline solution. The patient applied part of the probe is immersed into the saline
solution.) Measure current flowing in the circuit. = leakage current.
As a minimum, tests according to IEC606601-1 must be performed once a year. The requirements for
Body Floating (BF) have to be applied for TEE and Trans thorax probes bearing the symbol for safety
class BF.
The symbol for BF is indicated on the probe connector label below:
Figure 10-6 GE Probe Connector Label example
Where applicable, a typical test setup of non-TEE Probes can be as illustrated in:
Leakage Current" on page 10-25.
A typical test setup for TEE probes could be as indicated below:
The test passes when the reading measure less than the values in: Table 10-11 on page 10-15.
WARNING
WARNING
The handle of the TEE probes must not be immersed.
Содержание LOGIQ E10
Страница 1: ...LOGIQ E10 Basic Service Manual 5750007 1EN Rev 1 Version R1 2018 GENERAL ELECTRIC COMPANY...
Страница 2: ......
Страница 3: ...DIRECTION 5750007 1EN REV 1 LOGIQ E10 BASIC SERVICE MANUAL i Important Precautions TRANSLATION POLICY...
Страница 4: ...DIRECTION 5750007 1EN REVISION REV 1 LOGIQ E10 BASIC SERVICE MANUAL ii...
Страница 5: ...DIRECTION 5750007 1EN REV 1 LOGIQ E10 BASIC SERVICE MANUAL iii...
Страница 6: ...DIRECTION 5750007 1EN REVISION REV 1 LOGIQ E10 BASIC SERVICE MANUAL iv...
Страница 7: ...DIRECTION 5750007 1EN REV 1 LOGIQ E10 BASIC SERVICE MANUAL v...
Страница 8: ...DIRECTION 5750007 1EN REVISION REV 1 LOGIQ E10 BASIC SERVICE MANUAL vi...
Страница 9: ...DIRECTION 5750007 1EN REV 1 LOGIQ E10 BASIC SERVICE MANUAL vii...
Страница 10: ...DIRECTION 5750007 1EN REVISION REV 1 LOGIQ E10 BASIC SERVICE MANUAL viii...
Страница 11: ...DIRECTION 5750007 1EN REV 1 LOGIQ E10 BASIC SERVICE MANUAL ix...
Страница 12: ...DIRECTION 5750007 1EN REVISION REV 1 LOGIQ E10 BASIC SERVICE MANUAL x...
Страница 13: ...DIRECTION 5750007 1EN REV 1 LOGIQ E10 BASIC SERVICE MANUAL xi...
Страница 14: ...DIRECTION 5750007 1EN REVISION REV 1 LOGIQ E10 BASIC SERVICE MANUAL xii...
Страница 18: ...DIRECTION 5750007 1EN REVISION REV 1 LOGIQ E10 BASIC SERVICE MANUAL xvi This page was intentionally left blank...
Страница 34: ...DIRECTION 5750007 1EN REV 1 LOGIQ E10 BASIC SERVICE MANUAL xxxii Hardware Kit 9 49...
Страница 776: ...DIRECTION 5750007 1EN REV 1 LOGIQ E10 BASIC SERVICE MANUAL Index 4...
Страница 777: ......
Страница 778: ...2018 General Electric Company GE Healthcare 9900 Innovation Drive RP 2156 Wauwatosa WI 53226 USA www gehealthcare com...