September 2007
18
HygroTrace Moisture Transmitter
User
To specify an offset value in PPB
and add to the measured PPB value,
or to dry down the sensor, select
User...
on the Main Menu, press
Enter
and see the following screen:
Use the arrow keys to enter the user
passcode (2719), press
Enter
and the
following screen appears. The
User
Mode
indicator
U
appears in the
lower right corner of the screen.
PPB Offset
To specify a PPB offset value, select
PPB Offset
, press
Enter
and the
following screen appears:
Use the arrow keys to enter the PPB
offset desired, press
Enter
and the
screen returns to the
User Settings
Menu
.
Clear PPB Offset
To clear the PPB offset entered
above, select
Clear PPB Offset
and
press
Enter
. The following screen
appears. Press
Cancel
to return to
the
Main Menu
or proceed to the
next option.
Select
YES
or
NO
and press
Enter
, or
press
Cancel
. If you select
NO
or
press
Cancel
, the unit returns to the
previous screen and the PPD offset
is not cleared.
If you select
YES
, the
PPB Offset
is cleared and the unit returns to
the previous screen.