Section 6-2 - Input AC Voltage Configuration
Section 6-2
Input AC Voltage Configuration
•
The internal Vivid™ 4
voltage is 230V AC.
•
The Back End and Front End Crates operate on 230V AC.
•
The Peripherals operate on local voltage.
•
The Monitor operates on 110V AC internal.
6-2-1
Secondary Voltage Configuration
The AC input voltage is factory
preset according to specific requirements, as indicated on the label near
the AC input connector. The AC transformer’s primary and secondary voltages are set up by using the
appropriate plugs connected in Jumper1 and Jumper2 on the AC Distribution Box, in accordance with
the information in
6-2-2
AC Input Cord
The input voltage power cord and plug must meet the local standards and requirements.
Table 6-2 Primary and Secondary AC Voltages
Jumper 1
PRIMARY
Voltage
Range
100-120V
220-240V
Plug P/N
2253054
2253056
Jumper 2
SECONDARY
Voltage
100V
120V
220-230V
220-240V
Plug P/N
2253057
2253058
2253059-2
2253060-2