Appendix E: Installing Wall Mount Units
Wall Mount Bracket Installation
Service Manual
E-5
Note:
When a wall mount bracket is replaced, make sure to transfer the warmer unit serial number on to
the blank label provided on the new bracket.
Note:
The screws provided (3/8 - 16 UNC, 2 1/2” long) are for installations in 1/2” or 5/8” thick drywall. For
applications involving double thicknesses of drywall, longer screws will be required. To properly
mount the warmer, at least 2 screw threads must be engaged.
Call out
Description
Orderable Service Part Number
1
Toggler anchor bolt
6600-0413-400
*
* Replacement parts are also orderable through Bracket FRU Kit and Bumper FRU Kit.
Refer to
for details.
2
Flat washer
6600-0173-400
*
3
Split lock washer
6600-0440-400
*
4
Hex head screw
6600-0414-400
*
5
M6 screw with captive washers
M1110521
FIGURE E-4. Wall Mounting
WALL MOUNT
BRACKET
7"
(17.8 cm)
WALL BUMPER
11"
(28 cm)
WALL MOUNT WARMER SYSTEM
8.5" O.C.
(21.5 CM)
TRACK
STUD
DRY WALL
ø 3/4" HOLE
(ø 1.9 cm)
Содержание Giraffe
Страница 4: ...RH 2 Service Manual ...
Страница 12: ...ii Service Manual About this Manual User Responsibility ...
Страница 38: ...Chapter 1 Functional Description System Functions 1 14 Service Manual FIGURE 1 8 Power Failure ...
Страница 44: ...Chapter 1 Functional Description System Functions 1 20 Service Manual FIGURE 1 14 Manual Mode Heat Control ...
Страница 60: ...Chapter 2 Installation Maintenance and Checkout Scale Checkout Procedures 2 8 Service Manual ...
Страница 62: ...Chapter 3 Calibration Scale Calibration Bedded Models Only 3 2 Service Manual ...
Страница 172: ...Chapter 6 Illustrated Parts Wiring Diagrams 6 40 Service Manual ...
Страница 180: ...Appendix A Specifications SpO2 Specifications A 8 Service Manual ...
Страница 196: ...Appendix D Troubleshooting Service Mode Application Software 1 0 9 and Earlier D 8 Service Manual ...