Appendix D: Troubleshooting
Service Mode (Application Software 1.0.9 and Earlier)
Service Manual
D-7
Byte 5 – Manual Temp Alarm Setting.
“SPO2SET” With
Masimo
SpO
2
installed:
Byte 0 – The Pulse Tone Volume.
Byte 1 – Averaging (0x0 for “2-4”, 0x02 for “8”, and 0x06 for “16”).
Byte 2 – Sensitivity (0x0 for Maximum, 0x01 for Normal).
Byte 3 – Low Pulse Rate Limit Setting.
Byte 4 – High Pulse Rate Limit Setting.
Byte 5 – Low Saturation Limit Setting.
Byte 6 – High Saturation Limit Setting.
With Nellcor SpO
2
installed:
Byte 0 – The Pulse Tone Volume.
Byte 1 – Response Mode (0x01 for Normal, 0x02 for Fast).
Byte 2 – SatSeconds™ Limit Setting.
Byte 3 – Low Pulse Rate Limit Setting.
Byte 4 – High Pulse Rate Limit Setting.
Byte 5 – Low Saturation Limit Setting.
Byte 6 – High Saturation Limit Setting.
Содержание Giraffe
Страница 4: ...RH 2 Service Manual ...
Страница 12: ...ii Service Manual About this Manual User Responsibility ...
Страница 38: ...Chapter 1 Functional Description System Functions 1 14 Service Manual FIGURE 1 8 Power Failure ...
Страница 44: ...Chapter 1 Functional Description System Functions 1 20 Service Manual FIGURE 1 14 Manual Mode Heat Control ...
Страница 60: ...Chapter 2 Installation Maintenance and Checkout Scale Checkout Procedures 2 8 Service Manual ...
Страница 62: ...Chapter 3 Calibration Scale Calibration Bedded Models Only 3 2 Service Manual ...
Страница 172: ...Chapter 6 Illustrated Parts Wiring Diagrams 6 40 Service Manual ...
Страница 180: ...Appendix A Specifications SpO2 Specifications A 8 Service Manual ...
Страница 196: ...Appendix D Troubleshooting Service Mode Application Software 1 0 9 and Earlier D 8 Service Manual ...