Appendix C: Alarm Priorities
C-2
Service Manual
7
Baby
temperature
>2.0° from
Control Temp
Baby Cold-
Check
Temp
Probe
5 min
1
With hysteresis to
avoid nuisance
alarm recurrence
after silencing.
Hysteresis set at
0.2°C - alarm
clears at 1.8°C
Primary
Operational in
Baby Control
Mode
8
Check baby
Radiant %
power @ 100%
for > 12 min
Check
Baby
12 min
2
Primary
Operational in
Baby Control
Mode
9
Check baby
Radiant Heater
% above
“Check Baby
Alarm
Disabled”
(CPAD) limit for
> 12 min
Check
Baby
12 min
2
Primary
Operational in
Baby Control
Mode
10
Temperature
probe failure
Two
Thermistors in
a probed differ
by 0.5° C or
more.
Temp.
Probe
Failure
2 min
2 to 1
(after 30 secs.)
Primary
Operational in
Baby Control
mode or
Manual Mode
when Manual
Temperature
alarm set
11
Disconnected
temperature
probe
No longer
getting a
reading from
the
temperature
probe
Confirm
Probe Jack
Connection
2 min
2 to 1
(after 30 secs.)
Primary
Operational in
Baby Control
mode or
Manual Mode
when Manual
Temperature
alarm set
12
Baby
temperature
>1.0°C (>0.5°C)
from control
temp
Baby Hot -
Check
Temp
Probe
12 min
2
With hysteresis to
avoid nuisance
alarm recurrence
after silencing.
Hysteresis set at
0.2°C - alarm
clears at 0.8°C
(0.3°C)
Primary
Operational in
Baby Control
Mode
13
Baby
temperature
<1.0°C (<0.5°C)
from control
temp
Baby Cold -
Check
Temp
Probe
12 min
2
With hysteresis to
avoid nuisance
alarm recurrence
after silencing.
Hysteresis set at
0.2°C - alarm
clears at 0.8°C
(0.3°C)
Primary
Operational in
Baby Control
Mode
Priority Alarm Subject
Activation
Criteria
Message
Displayed
Silence
Time
Sound:
1 = High
2 = Low
3 = informational
Notes/
Comments
Indication
Light
Special
Requirements
Содержание Giraffe
Страница 4: ...RH 2 Service Manual ...
Страница 12: ...ii Service Manual About this Manual User Responsibility ...
Страница 38: ...Chapter 1 Functional Description System Functions 1 14 Service Manual FIGURE 1 8 Power Failure ...
Страница 44: ...Chapter 1 Functional Description System Functions 1 20 Service Manual FIGURE 1 14 Manual Mode Heat Control ...
Страница 60: ...Chapter 2 Installation Maintenance and Checkout Scale Checkout Procedures 2 8 Service Manual ...
Страница 62: ...Chapter 3 Calibration Scale Calibration Bedded Models Only 3 2 Service Manual ...
Страница 172: ...Chapter 6 Illustrated Parts Wiring Diagrams 6 40 Service Manual ...
Страница 180: ...Appendix A Specifications SpO2 Specifications A 8 Service Manual ...
Страница 196: ...Appendix D Troubleshooting Service Mode Application Software 1 0 9 and Earlier D 8 Service Manual ...