7.4.2
Mounting the System
WARNING:
To avoid patient injury, do not mount the system while a patient occupies the bed.
1. Make sure the light pipe is securely attached to the system as instructed in section 7.4.1.
2. Use the provided hex key to loosen the two mounting screws on the side of the dovetail mount
bracket (Refer to Figure 7-3).
3. Position the system on the accessory dovetail rail.
4. Secure the system to the dovetail rail by using the hex key to tighten the two mounting screws (Refer
Note: If you intend to plug the power cord into an accessory outlet on a microenvironment, refer to the
microenvironment manual for proper power cord connection.
5. Plug the power cord into the power inlet connector and rotate the retainer to grip and secure the
power cord and then plug it into the power outlet.
NOTE:
If you intend to plug the power cord into an accessory outlet on a microenvironment, refer to
the microenvironment manual for proper power cord connection.
6. Before use, follow the checkout procedures as described in Chapter 8.
Figure 7-3
Figure 7-4
49
Chapter 7: Installation
Содержание Giraffe Blue Spot PT Lite
Страница 1: ...Giraffe Blue Spot PT Lite Operation Maintenance and Service Manual GE Healthcare ...
Страница 7: ... 7 Part I Operation and Maintenance ...
Страница 8: ...8 Part I Operation and Maintenance This page is intentionally left blank ...
Страница 16: ...This page is intentionally left blank 16 Chapter 1 Product Overview ...
Страница 22: ...This page is intentionally left blank 22 Chapter 2 Product Setup and Operation ...
Страница 26: ...This page intentionally left blank 26 Chapter 3 Operator s Maintenance ...
Страница 32: ...This page intentionally left blank 32 ...
Страница 33: ... 33 Part II Service ...
Страница 34: ...34 Part II Service This page is intentionally left blank ...
Страница 44: ...44 Important Service Safety Information This page is intentionally left blank ...
Страница 50: ...This page is intentionally left blank 50 Chapter 7 Installation ...
Страница 55: ...Chapter 9 Calibration No calibration is required for the Giraffe Blue Spot PT Lite 55 ...
Страница 56: ...This page is intentionally left blank 56 Chapter 9 Calibration ...
Страница 72: ...This page is intentionally left blank 72 Chapter 11 Replacement Procedures ...
Страница 88: ...12 5 Wiring Diagram 88 Chapter 12 Service Parts ...
Страница 94: ...This page intentionally left blank 94 ...
Страница 95: ......