6 Select the instruction
AccumulatorWash
in the Frac list.
7 Select the number of strokes to be used for washing with the parameter
Strokes
.
8 Click on the
Execute
button.
Setting delay volume
The delay volume between the UV flow cell in the chromatographic system and Fraction
Collector Frac-950 must be known to UNICORN. The delay volume is used to adapt the
collected fractions to the event marks generated by UNICORN.
1 In UNICORN, select menu
System Control:System:Settings
2 Click the
Specials
radio button and select instruction
FracParameters
. The
DelayVol
instruction becomes highlighted.
3 To change the setting, click on the up and down arrows for the
DelayVol
parameter,
or type a new value directly in the parameter window.
4 The appropriate value for your ÄKTA system is found in the
ÄKTA design Optional
Configuration User Manual
.
5 Click on the
OK
button. The value entered will be used until a further change is made.
44
Frac-920 and Frac-950 Operating Instructions 28-9622-09 AC
4 Operation
4.3 Fraction Collector Frac-950