 [EN]  English - 3

K346 Issue 2

To start - Power on or off

To turn the instrument on or off, press 

 (A1 - item [1]). The 

instrument does a self test and then shows the applicable 
data.
When the power is off, the last set of configuration options 
stays in memory. Refer to “Maintenance”.

To start - Set up the basic operation

Use the 

Set Up

 menu to set up the basic operation of the 

instrument. 

If there is additional data for a menu option, select 

Settings

(

■ ■

) to see the values that are set up. If 

necessary, adjust the values.

Table 1: 

(

Part of table) Menu options - Set Up

Table 1: (Part of table) Menu options - Set Up 

To start - Select a task (Measure and/or simulate)

When the instrument is set up (Table 1), use the 

Select Task

 

menu to select the applicable task.

In Table 2/3, IDOS is a Universal Measurement Module 
(UMM). If you attach a UMM to the communications port 
(A1 - item [9]), the 

Select Task

 menu shows the applicable 

IDOS options.

Table 2: Menu options - Select Task

Table 3 shows all the one and two function operations that 
are available. If you attach a UMM, you can only use the 
options that include IDOS.

1

Menu:

Select Task

2

3

Menu:

Set Up

4

5

(Table 2)

(Table 1)

[

]/[  ]

Options

(If applicable)

Description

DPI 822 only. To add a series resistor into the mA 
circuit. You can then use this instrument together 
with a HART

®

 communicator to set up and 

calibrate HART

®

 devices.

... Scale To select the applicable international temperature 

scale: IPTS 68 or ITS 90.

CJ ... To select the type of cold junction (CJ) 

compensation.

Automatic

: The instrument monitors the CJ 

temperature and applies the necessary CJ 
compensation.

Manual

: Measure the CJ temperature and set the 

applicable value. The instrument uses this value to 
apply the necessary CJ compensation.

Additional data (Manual): Select Settings 

(

 

)

To select and set up the backlight fa timer.

Additional data: Select Settings 

(

 

)

To select and set up the power off fa timer. 

Additional data: Select Settings 

(

 

)

To show the battery level (%).

To set the display contrast (%).

▲ 

Increases %, 

▼ 

decreases %

Options

(If applicable)

Description

To set the time + date. The calibration facility uses 
the date to give service and calibration messages. 
To set the language option.

To calibrate the instrument.

Additional data: Refer to “Calibration”.

To select and show the applicable status data. 
(Software Build, Calibration Due date, Serial 
Number, IDOS Information).

1

Menu:

Select Task

2

3

Display:

T1 output 

(Table 2/3)

Sk1 = Edit
Sk2 = Settings

Options

(If applicable)

Description

 T1 

or

mV

An input measurement task: 
T1 - Measure thermocouple temperature  OR
mV - Measure thermocouple mV.

T1 or
mV

An output task: 
T1 - Simulate thermocouple temperature  OR
mV - Simulate thermocouple mV.

mA

DPI 822 only. A mA measurement task. 

mA(24V)

DPI 822 only. A mA measurement task + the loop 
power supply is on. 
DPI 822 only. A switch test. 

 

IDOS

UMM only. An IDOS measurement task. 
UPM

 

only. A leak test. 

To set up the way the instrument works.

Additional data: Refer to: Set Up 

(Table 1).

Содержание Druck DPI 821

Страница 1: ...Thermocouple calibrator and Thermocouple loop calibrator User manual K346 GE Infrastructure Sensing 99 Washington Street Melrose MA 02176 Phone 781 665 1400 Toll Free 1 800 517 8431 Visit us at www TestEquipmentDepot com ...

Страница 2: ...K346 Issue 2 A1 A2 DPI 822 A3 10 1 3 4 2 5 6 7 9 8 A2 A3 11 B1 12 14 15 13 20 21 22 25 17 18 26 19 27 16 23 24 B1 Customer service Visit our web site www gesensing com ...

Страница 3: ...mes are trademarks of their respective companies Introduction The DPI 821 Thermocouple Calibrator and DPI 822 Thermocouple Loop Calibrator are part of the Druck DPI 800 series of hand held instruments The DPI 800 series uses Intelligent Digital Output Sensor IDOS technology to give instant plug and play functionality with a range of Universal Measurement Modules UMM Example the Universal Pressure ...

Страница 4: ...Selects or accepts an item or value Selects or cancels a selection 7 Right hand soft key Selects the function above it on the display Item 25 Example Settings 8 Display Refer to A3 9 SENSOR PC Communications port Use to connect a Universal Measurement Module UMM or a RS232 cable 10 Thermocouple connector Refer to Operation 11 Connection point for some of the optional accessories Refer to the datas...

Страница 5: ...applicable international temperature scale IPTS 68 or ITS 90 CJ To select the type of cold junction CJ compensation Automatic The instrument monitors the CJ temperature and applies the necessary CJ compensation Manual Measure the CJ temperature and set the applicable value The instrument uses this value to apply the necessary CJ compensation Additional data Manual Select Settings To select and set...

Страница 6: ... type K J T To include maximum minimum and average values for the measurement task To select and set up a tare value for the measurement task a specified value or the reading on the display The instrument subtracts a positive tare value and adds a negative tare value Additional data Select Settings To select and set up a scale of values One local scale for each measurement task Maximum 5 Additiona...

Страница 7: ... If applicable Description Step To select and set up values for the Step output Example output cycle Auto Repeat Optional Additional data Table 6 Select Settings Step To select and set up values for the Defined Step output Example output cycle Auto Repeat Optional Additional data Table 6 Select Settings To select and set up values for the Ramp output Example output cycle Auto Repeat Optional Addit...

Страница 8: ...ure thermocouple values To measure the temperature or mV values of a thermocouple 1 Connect the instrument Figure 1 and if necessary adjust the Set Up Table 1 2 Select a T1 or mV input task from Select Task Table 2 3 and if necessary adjust the Settings Table 4 Figure 1 Example configuration To measure the temperature or mV values of a thermocouple Operation Change the output values When the outpu...

Страница 9: ...mocouple Operation Transmitter calibration DPI 822 only To calibrate a transmitter 1 Connect the instrument Figure 3 4 and if necessary adjust the Set Up Table 1 2 Select the applicable calibration task from Select Task Table 2 3 and if necessary adjust the Settings Table 4 5 3 Supply the output values to the system Table 7 Figure 3 Example configuration Transmitter calibration with external loop ...

Страница 10: ...losed in the top right hand corner 3 Supply the output values to the system Table 7 Example Nudge output a Use Edit to set a value less than the switch value b Use the buttons to change the value in small increments Example Ramp output a Set High and Low values that are applicable to the switch value Table 6 Then to get an accurate switch value set a long Travel period b Use Start Stop to start an...

Страница 11: ...he Settings Table 4 5 2 If necessary do a zero correction Table 4 To measurepressure withanother operation Figure 8 use the same procedure Figure 8 Example configuration To measure pressure and temperature UPM Leak test To do a leak test on a pressure system Figure 7 1 Select an applicable leak test from Select Task Table 2 3 and if necessary adjust the Set Up Table 1 and the Settings Table 4 2 Se...

Страница 12: ...the display shows the applicable instructions to complete the calibration When the calibration is complete select Calibration Due and set the new calibration date for the instrument Calibration Procedures mV input or output 1 Connect the instrument to the calibration equipment mV input Figure 1 mV output Figure 2 2 Let the equipment get to a stable temperature minimum 5 minutes since the last powe...

Страница 13: ...re that the calibration is correct select the applicable mA input task Table 2 and apply these values mA 55 40 24 18 12 6 0 short circuit Then mA 0 6 12 18 24 40 55 5 Make sure that the error is in the specified limits Table 11 Table 11 mA input error limits Calibration Procedures IDOS UMM Refer to the user manual for the IDOS UMM When the calibration is complete the instrument automatically sets ...

Страница 14: ...38 1832 150 1000 0 7 0 4 N IEC 584 454 4 270 20 1 8 1 0 N IEC 584 4 2372 20 1300 1 1 0 6 L DIN 43710 328 1652 200 900 0 6 0 3 U DIN 43710 328 212 200 100 0 9 0 5 U DIN 43710 212 1112 100 600 0 6 0 3 C 32 2732 0 1500 1 8 1 0 C 2732 3632 1500 2000 2 5 1 4 C 3632 4208 2000 2320 3 4 1 9 D 32 3092 0 1700 1 8 1 0 D 3092 3992 1700 2200 2 9 1 6 D 3992 4523 2200 2495 6 5 3 6 Cold Junction CJ error Maximum ...

Отзывы: