Routine Maintenance
© 2015 General Electric Company – All rights reserved.
ADTS 405 Mk2 Instruction Manual-English | 55
Figure 4-3: ADTS 405F DC Fuse Positioning
4.4.3.3 Replacing ADTS 405R AC Fuses
On the ADTS 405R, two 20mm High Breaking Capacity fuses are located in the holder mounted on
the power input connector on the rear of the rack and protect the unit.
a.
Set the power switch to OFF.
b.
Isolate the power supply to the instrument and disconnect the IEC power supply
a.
connector.
b.
Pull the capped lid of the fuse carrier (2) on the power supply input socket assembly.
c.
Replace the fuse cartridges (3).
d.
Refit the fuse carrier (2) flush in the power supply inlet socket assembly.
e.
Refit and reconnect rack-mounted units as detailed in Section 2, INSTALLATION.
f.
Switch on the power supply and set the power switch to ON. The instrument should now
be operational.
Note:
(1) To gain access to rack-mounted instruments, it may be necessary to
partially or completely withdraw the instrument. Isolate pneumatic supplies
and depressurise all pressure supply inlet and outlet lines.
Note:
(2) If the fuse blows immediately after switch-on, contact the
manufacturer or Service Agent.
1
3
2
Содержание Druck ADTS 405 R/F Mk2
Страница 2: ......
Страница 10: ...viii ADTS 405 Mk2 Instruction Manual English 2015 General Electric Company All rights reserved ...
Страница 14: ...xii ADTS 405 Mk2 Instruction Manual English 2015 General Electric Company All rights reserved ...
Страница 16: ...xiv ADTS 405 Mk2 Instruction Manual English 2015 General Electric Company All rights reserved ...
Страница 18: ...xvi ADTS 405 Mk2 Instruction Manual English 2015 General Electric Company All rights reserved ...
Страница 121: ......