Testing Aircraft Systems or UUT
© 2015 General Electric Company – All rights reserved.
ADTS 405 Mk2 Instruction Manual-English | 37
3.6.2.1 To change units in FULL SETUP Mode as follows:
a.
Press
SETUP
.
b.
Select [UNITS].
c.
Select required pressure units for display using [NEXT], [PREV] and [SAVE].
d.
Use
CLEAR/QUIT
to return to pressure display.
3.6.2.2 To change units in MINIMUM SETUP Mode as follows:
a.
Press
SETUP
.
b.
Select [UNITS TYPE].
c.
Select [PRESS].
d.
Use
CLEAR/QUIT
to return to the main pressure display.
Note:
Pressure units cannot be selected when set-up mode is switched off.
Holding down the F1 key while pressing SETUP provides access to the CONFIGuration mode.
CONFIGuration is similar to SETUP with many identical functions. The changes made under
CONFIGuration are permanent and remain set after the system is switched off. Configuration can
be used to change the power-up default settings of the ADTS 405 functions:
a.
Holding down F1 then pressing SETUP causes the system to enter CONFIGuration mode.
CONFIGuration may be password protected using a four digit PIN. If PIN protected, the
display prompts for entry of the PIN. Press each digit of the PIN in turn (ENTER is not
required) as each digit is entered, an asterisk appears. If an incorrect PIN is entered, the
complete PIN is requested again.
b.
When the correct PIN is entered, the CONFIG menu appears.
c.
The PIN may be changed using CONFIG,-000.
d.
To use the configuration menu, press the appropriate function key to select the required
menu item. To configure a parameter key, enter configuration menu then press the
required parameter key. Each parameter key configuration provides changes to the normal
user mode function and to the settings available in the set-up menu.
e.
Once a parameter has been changed in configuration, use
CLEAR/QUIT
to return to the
previous menu. After all required key parameters have been changed, repeatedly press
CLEAR/QUIT
until the display shows the normal operating mode.
Содержание Druck ADTS 405 R/F Mk2
Страница 2: ......
Страница 10: ...viii ADTS 405 Mk2 Instruction Manual English 2015 General Electric Company All rights reserved ...
Страница 14: ...xii ADTS 405 Mk2 Instruction Manual English 2015 General Electric Company All rights reserved ...
Страница 16: ...xiv ADTS 405 Mk2 Instruction Manual English 2015 General Electric Company All rights reserved ...
Страница 18: ...xvi ADTS 405 Mk2 Instruction Manual English 2015 General Electric Company All rights reserved ...
Страница 121: ......