Chapter 4. MAINTENANCE
54 | ADTS 405 Mk2 Instruction Manual-English
© 2015 General Electric Company – All rights reserved.
4.4.3.1 Replacing ADTS 405F AC Fuses
On the ADTS 405F two 20mm High Breaking Capacity fuses are located in the holders mounted on
the pump rack front panel and protect the entire unit.
Figure 4-2: ADTS 405F AC Fuse Positioning
a.
Set the power switch to OFF.
b.
Isolate the power supply to the instrument and disconnect the power supply connector (1).
c.
Remove the fuse holders (2) from the front panel and replace the fuse cartridges (3).
d.
Refit the fuse holders (2) in the front panel.
e.
Connect the power supply connector and switch on the power supply.
f.
Set the unit power switch to ON.
Note:
If the fuse blows immediately after switch-on contact the manufacturer
or Service Agent.
4.4.3.2 Replacing ADTS 405F DC Fuses
On the ADTS 405F, two 1.25” Time Lag fuses are located in the holders mounted on the pump rack
front panel and protect the entire unit. The T20AH 250V fuse (left) is connected in the positive (+)
feed to the pressure controller rack. The T32A fuse (right) is connected in the common (0v) return
from both the Controller rack and Pump rack.
Note:
On the ADTS 405R, DC over-current protection fusing is internal and not
user serviceable.
1
2
3
Содержание Druck ADTS 405 R/F Mk2
Страница 2: ......
Страница 10: ...viii ADTS 405 Mk2 Instruction Manual English 2015 General Electric Company All rights reserved ...
Страница 14: ...xii ADTS 405 Mk2 Instruction Manual English 2015 General Electric Company All rights reserved ...
Страница 16: ...xiv ADTS 405 Mk2 Instruction Manual English 2015 General Electric Company All rights reserved ...
Страница 18: ...xvi ADTS 405 Mk2 Instruction Manual English 2015 General Electric Company All rights reserved ...
Страница 121: ......