DigitalFlow™ XMT868i Startup Guide
67
Appendix A. CE Mark Compliance
Appendix A.
CE Mark Compliance
A.1 Introduction
For CE Mark compliance, the Model XMT868i flow transmitter must be wired in accordance with the instructions in
this appendix.
IMPORTANT:
CE Mark compliance is required for all units intended for use in EU countries.
A.2 Wiring
The Model XMT868i must be wired with the recommended cable, and all connections must be properly shielded and
grounded. Refer to
below for the specific requirements.
Note:
If the Model XMT868i is wired as described in this appendix, the unit will comply with the EMC Directive.
Table 21: Wiring Requirements
Connection
Cable Type
Ground Termination
Transducer
Armored RG62 a/U
Grounded using a cable gland
Input/Output Armored 22 AWG, shielded with armored material added to
outside of jacket
Grounded using a cable gland
Power
Armored 14 AWG, 3 conductor
Grounded using a cable gland
Содержание DigitalFlow XMT868i
Страница 2: ......
Страница 4: ...ii no content intended for this page ...
Страница 28: ...Chapter 1 Installation 20 DigitalFlow XMT868i Startup Guide no content intended for this page ...
Страница 34: ...Chapter 1 Installation DigitalFlow XMT868i Startup Guide 26 no content intended for this page ...
Страница 52: ...Chapter 2 Initial Setup 44 DigitalFlow XMT868i Startup Guide no content intended for this page ...
Страница 76: ...Appendix A CE Mark Compliance 68 DigitalFlow XMT868i Startup Guide no content intended for this page ...
Страница 86: ...Warranty 78 DigitalFlow XMT868i Startup Guide no content intended for this page ...
Страница 90: ...Index 82 DigitalFlow XMT868i Startup Guide no content intended for this page ...
Страница 92: ... no content intended for this page ...
Страница 94: ... no content intended for this page ...
Страница 95: ......