DigitalFlow™ XMT868i Startup Guide
29
Chapter 2. Initial Setup
2.2 XMT868i Enclosure Keypad Program (cont.)
The meter then starts to display measured parameters.
To enter the
Keypad Program
, press the
[Escape]
key, followed by the
[Enter]
key, and the
[Escape]
key again. Each
successive key must be entered within 10 seconds of the prior key.
As a guide in following the programming instructions in this chapter, the relevant portions of the Model XMT868i
menu map have been reproduced in
and
Proceed to the following sections
to enter data in the
Channel
or
GLOBL
menus.
IMPORTANT:
If the keypad has not been pressed for 10 minutes, the XMT868i exits the
Keypad Program
and returns to
displaying measurements. The meter retains any configuration changes that were confirmed with the
[Enter]
key, and restarts as if the operator had completed the programming cycle.
2.3 Entering Data in the Global Menu
To begin programming your meter, you must select the system units from the
GLOBL
menu as discussed below. Refer
to
and remember to record all programming data in Appendix B,
Data Records
.
Note:
Refer to the
Programming Manual
for information on the other submenus in the
GLOBL
menu.
2.3.1 Entering Global System Data
The
GLOBL-SYSTM
submenu is used to enter several general system parameters (i.e., English or metric units). For
units with 2 channels, this menu is also used to compute parameters such as the sum, difference or average of the
channel 1 and channel 2 signals. When calculating the
SUM
,
DIF
or
AVE
readouts, data from the
GLOBL-SYSTM
submenu is used. Any conflicting data entered in the
Channel Systm
submenu is overridden.
1.
In the
Keypad Program
, scroll to
PROG
and press
[Enter]
.
2.
In the
PROG
menu, scroll to
GLOBL
and press
[Enter]
.
3.
In the
Global PROGRAM
menu, scroll to
SYSTM
and press
[Enter]
.
4.
The first prompt asks for a
Meter Message.
Enter a numeric or text message (up to 16 characters) and press
[Enter]
.
5.
Scroll to the desired
System Units
selection (either metric or English) and press
[Enter]
.
The XMT868i will
display all parameters and measurements in the designated units.
•
For a single-channel XMT868i, the program returns to the
Global PROGRAM
menu. Scroll to CH1 or CH2 to enter
the
Channel
menu discussed on
•
For a 2-channel meter, proceed to step 6.
CH1 VEL E1
10.00 Ft/s
Содержание DigitalFlow XMT868i
Страница 2: ......
Страница 4: ...ii no content intended for this page ...
Страница 28: ...Chapter 1 Installation 20 DigitalFlow XMT868i Startup Guide no content intended for this page ...
Страница 34: ...Chapter 1 Installation DigitalFlow XMT868i Startup Guide 26 no content intended for this page ...
Страница 52: ...Chapter 2 Initial Setup 44 DigitalFlow XMT868i Startup Guide no content intended for this page ...
Страница 76: ...Appendix A CE Mark Compliance 68 DigitalFlow XMT868i Startup Guide no content intended for this page ...
Страница 86: ...Warranty 78 DigitalFlow XMT868i Startup Guide no content intended for this page ...
Страница 90: ...Index 82 DigitalFlow XMT868i Startup Guide no content intended for this page ...
Страница 92: ... no content intended for this page ...
Страница 94: ... no content intended for this page ...
Страница 95: ......