April 2008
1-8
Calibration
aLOW Option = [F1]
1.
Enter the low reference value and press the
[ENT]
key.
2.
Press
[F1]
to store the current low reference value or press
[F2]
to
cancel the entry. In either case, the
ANALOG INPUT
prompt will
reappear.
3.
Press
[F2]
to proceed to the next section.
aHIGH Option = [F2]
1.
Enter the high reference value and press the
[ENT]
key.
2.
Press
[F1]
to store the current high reference value or press
[F2]
to
cancel the entry. In either case, the
ANALOG INPUT
prompt will
reappear.
3.
Proceed to
Procedure Options
below.
4 mA Option = [F3]
1.
Connect the 4 mA current source to the currently selected analog
input, as shown in Figure 1-3 on page 1-6.
2.
Press
[F1]
to store the current 4 mA value or press
[F2]
to cancel the
entry. In either case, the
ANALOG INPUT
prompt reappears.
3.
Press
[F4]
to proceed to the next section.
20 mA Option = [F4]
1.
Connect the 20 mA current source to the currently selected analog
input, as shown in Figure 1-3 on page 1-6.
2.
Press
[F1]
to store the current 20 mA value or press
[F2]
to cancel
the entry. In either case, the
ANALOG INPUT
prompt reappears.
3.
Proceed to
Procedure Options
below.
Procedure Options
You have completed calibration of the analog output. Do one of the
following:
•
To calibrate another input, return to Step 3 in
Accessing the
Calibration Menu
on page 1-7.
•
To calibrate/test additional slots, press
[F4] (EXIT)
and proceed to
the appropriate section.
•
To leave the Calibration Menu, press
[F4] (EXIT)
twice.
Содержание DigitalFlow GC868
Страница 6: ...Chapter 1 ...
Страница 23: ...Chapter 2 ...
Страница 31: ...Chapter 3 ...
Страница 39: ...Chapter 4 ...
Страница 52: ...Appendix A ...
Страница 53: ...Service Record Introduction A 1 Data Entry A 1 Diagnostic Parameters A 3 ...
Страница 55: ...April 2008 A 2 Service Record Table A 1 Service Record Continued Date Description of Service Performed By ...
Страница 57: ...Appendix B ...