Calibration
1-7
April 2008
Calibrating the Analog
Inputs (cont.)
For this discussion, assume that the option card has been installed in
Slot x
.
Note:
The zero point of the analog input may be set for either 0 mA
or 4 mA. However, the calibration procedure always uses the 4
mA point, as the meter will extrapolate from this value to
obtain the 0 mA point.
Accessing the Calibration
Menu
1.
Press the
[CAL]
key to enter the
Calibration Program
.
2.
Press
[Fx]
to calibrate the desired slot. (The option bar will include
a slot listing for each installed option card.)
3.
Press
[Fx]
to select the desired input.
The procedure for calibrating
Input A
is identical to that for calibrating
Input B
. However, when calibrating
Input B
, be sure a calibrated
current source is connected to the appropriate pins on the terminal
block. See Figure 1-3 on page 1-6 for the correct pin numbers. To
calibrate a 0/4-20 mA input card, see the section below.
Note:
To calibrate an RTD input card, refer to page 1-9.
4.
Press
[F1]-[F4]
to select the reference point to be calibrated.
Do one of the following:
•
If you selected
[F1] (aLOW)
, proceed to
aLOW Option
on the next
page.
•
If you selected
[F2] (aHIGH)
, proceed to
aHIGH Option
on the next
page.
•
If you selected
[F3] (4 mA)
, proceed to
4 mA Option
on the next
page.
•
If you selected
[F4] (20 mA),
proceed to
20 mA Option
on the next
page.
Содержание DigitalFlow GC868
Страница 6: ...Chapter 1 ...
Страница 23: ...Chapter 2 ...
Страница 31: ...Chapter 3 ...
Страница 39: ...Chapter 4 ...
Страница 52: ...Appendix A ...
Страница 53: ...Service Record Introduction A 1 Data Entry A 1 Diagnostic Parameters A 3 ...
Страница 55: ...April 2008 A 2 Service Record Table A 1 Service Record Continued Date Description of Service Performed By ...
Страница 57: ...Appendix B ...