April 2008
2-6
Error Codes and Screen Messages
Duplicate name,
Enter another.
The site file or log name is already in use. Enter a different name.
End Time must exceed
Start Time by 5 min.
This message appears when in the LOG menu. Enter an end time that is
at least five minutes later than the start time.
Header invalid
An option card error indicating a programming failure or a loss of
memory. Contact the factory.
Log Active, END only
hit any key
When in the LOG menu, this message indicates that the log is still
compiling data. Only the End Time may be edited.
Log DONE, to inspect
hit any key
When in the LOG menu, this message indicates that the log is complete.
Hit any key to display the log.
Log hasn’t started! hit key
When in the LOG menu, this message indicates that the log has not yet
started.
OK, settling
The measurement has not stabilized. Wait before taking a reading.
Outside limits, value rejected.
When in the CAL menu, this message indicates that the calibration of the
analog output is invalid. Hit any key to clear the message, and the GC868
will default to the last valid calibration. The message will also appear if
there is no ammeter connected to the analog output during calibration.
Overflow
The display value overflowed. Reduce the number of decimal digits or
change the units.
??P<L Enter L again.
The entered path length (P) is less than the axial dimension (L). Enter a
new value for L.
range is X.XX to X.XX
When in the
User Program
, this message indicates that the entered
number is unacceptable. Hit any key and enter a number within the
allowable range.
Review calibration
An option card error indicating a programming failure or a loss of
memory. Contact the factory.
Review parameters
An option card error indicating a programming failure or a loss of
memory. Contact the factory.
Starting time must exceed
current time
This message appears when in the LOG menu STD option. Enter a start
time that is later than the current time.
Write error.
An option card error indicating a programming failure or a loss of
memory. Contact the factory.
Table 2-1: Screen Messages (Continued)
Message
Meaning
Содержание DigitalFlow GC868
Страница 6: ...Chapter 1 ...
Страница 23: ...Chapter 2 ...
Страница 31: ...Chapter 3 ...
Страница 39: ...Chapter 4 ...
Страница 52: ...Appendix A ...
Страница 53: ...Service Record Introduction A 1 Data Entry A 1 Diagnostic Parameters A 3 ...
Страница 55: ...April 2008 A 2 Service Record Table A 1 Service Record Continued Date Description of Service Performed By ...