Guidelines for Successful
Pacemaker Monitoring:
•
Multi-vector pace detection is
determined by simultaneous
analysis of the displayed ECG
Lead and the Vector Lead.
•
Adequate pacemaker
detection is directly
dependant on the quality of
the ECG waveform. Proper
skin preparation and
electrode placement are
essential.
•
Ensure that the pace detection
mode is activated.
•
All detected pacemaker spikes
will appear upright, uniform
and white on the display
screen.
•
If the monitor is not
adequately detecting
pacemaker spikes —
as evidenced by heart rate
double counting, pacemaker
spikes not detected, alarms for
low heart rate or asystole —
it is recommended that you
change the electrode
placement to a recommended
configuration.
•
After changing electrode
placement, always
RELEARN
the ECG waveform.
If adjusting the electrode
placement does not resolve
the detection issue, change
the pace detection mode to
Pace 1 and
RELEARN
.
Recommended Alternative Lead Placements
Содержание DASH 3000 V5
Страница 30: ...TAB TRIM COVER NON TAB TRIM TAB TRIM COVER NON TAB TRIM ...
Страница 42: ...TAB TRIM COVER NON TAB TRIM TAB TRIM COVER NON TAB TRIM ...
Страница 76: ...TAB TRIM COVER NON TAB TRIM TAB TRIM COVER NON TAB TRIM ...
Страница 90: ...TAB TRIM COVER NON TAB TRIM TAB TRIM COVER NON TAB TRIM Optional Equipment ...
Страница 104: ...TAB TRIM COVER NON TAB TRIM TAB TRIM COVER NON TAB TRIM ...