53
5001A EN 20020315
Environmental requirements
Room temperature
20 °C - 30 °C
Maximum temperature change
0.5 °C/min.
Relative humidity
10 % - 80 %
Magnetic field
Less than 5 Gauss; compliant with
EN 61000-4-8, Level 5
Sunlight exposure
Not to be operated in full sunlight
Physical emissions
Noise emission (sound power level according to DIN 45635 Part.27)
• During scanning
max. 65 dB(A)
• Standby
max. 46 dB(A)
Heat emission (at max. throughput with 35 cm x 43 cm image plates)
• Europe
0.8 kWh
• US
2730 BTU/hr
Cassette buffer capacity
10 cassettes of mixed sizes, both in input and output buffer
Performance
Throughput for high resolution
80 plates/h (size dependent)
Throughput for standard resolution
95 plates/h (size dependent)
Grayscale resolution
Data acquisition
12 bits/pixel
Output to processor
12 bits/pixel
5001 A EN UM Centricity CR MP3510.fm Page 53 Friday, March 15, 2002 10:33 AM