16
5001A EN 20020315
The keypad
Special functions can be accessed via the keypad. The keypad features the
following keys:
Emergency
key
To give an image the status ‘emergency’ when it is
sent to the image processing station.
Erase key
To erase images without digitizing them.
This must be done if:
• an image plate has not been used for more than
3 days;
• an image plate has been exposed to an
exceptionally high X-ray dose.
Key-operator
key
To access advanced functions (‘key-operator
functions’).
Service key
To access service-level functions.
Reserved for trained service personnel.
Escape key
To quit the current function or exit a menu without
saving modifications.
Confirm key
In key-operator mode:
• to select a menu.
• to accept an entry in a menu and go back to
operator mode.
5001 A EN UM Centricity CR MP3510.fm Page 16 Friday, March 15, 2002 10:33 AM